Background and aim: Adenoma detection rate (ADR) is the most important quality indicator for screening colonoscopy, due to its association with colorectal cancer outcomes. As a result, a number of techniques and technologies have been proposed that have the potential to improve ADR. The aim of this study was to assess the potential impact of new-generation high-definition (HD) colonoscopy on ADR within the Bowel Cancer Screening Programme (BCSP).
Method: This was a retrospective single-center observational study in patients undergoing an index screening colonoscopy. The examination was performed with either standard-definition colonoscopes (Olympus Q240/Q260 series) or HD colonoscopes (Olympus HQ290 EVIS LUCERA ELITE system) with the primary outcome measures of ADR and mean adenoma per procedure (MAP) between the 2 groups.
Results: A total of 395 patients (60.5% male, mean age 66.8 years) underwent screening colonoscopy with 45% performed with HD colonoscopes. The cecal intubation rate was 97.5% on an intention-to-treat basis and ADR was 68.6%. ADR with standard-definition was 63.13%, compared with 75.71% with HD (P = .007). The MAP in the HD group was 2.1 (± 2.0), whereas in the standard-definition group it was 1.6 (± 1.8) (P = .01). There was no significant difference in withdrawal time between the 2 groups. In the multivariate regression model, only HD scopes (P = .03) and male sex (P = .04) independently influenced ADR.
Conclusion: Olympus H290 LUCERA ELITE HD colonoscopes improved adenoma detection within the moderate-risk population. A 12% improvement in ADR might be expected to increase significantly the protection afforded by colonoscopy against subsequent colorectal cancer mortality.
Keywords: Adenoma detection rate; Bowel cancer screening program; Colonoscopy; Colorectal cancer; High definition.
Copyright © 2016 Elsevier Inc. All rights reserved.