Objective: To examine the accuracy of fetal RHD genotype and RHD pseudogene determination in a multiethnical population.
Methods: Prospective study involving D-negative pregnant women. Cell-free DNA was extracted from 1 ml of maternal plasma by an automated system (MagNA Pure Compact, Roche) and real-time PCR was performed in triplicate targeting the RHD gene exons 5 and 7. Inconclusive samples underwent RHD pseudogene testing by real-time PCR analysis employing novel primers and probe.
Results: A positive result was observed in 128/185 (69.2%) samples and negative in 50 (27.0%). Umbilical cord blood phenotype confirmed all cases with a positive or negative PCR result. Seven (3.8%) cases were found inconclusive (exon 7 amplification only) and RHD pseudogene testing with both conventional and real-time PCR demonstrated a positive result in five of them, while two samples were also RHD pseudogene negative.
Conclusion: Real-time PCR targeting RHD exons 5 and 7 simultaneously in maternal plasma is an accurate method for the diagnosis of fetal D genotype in our population. The RHD pseudogene real-time PCR assay is feasible and is particularly useful in populations with a high prevalence of this allele.
Keywords: DNA/analysis; RhD Blood group system; genotyping techniques; prenatal diagnosis/methods; sensitivity and specificity.
© 2016 Wiley Periodicals, Inc.