Purpose: MRI allows radiation-free imaging of the head and neck area. However, implant-supported prostheses may severely impair image quality due to artefacts. Therefore, identification of preferable material compositions for implants and supported prostheses with little impact on MR image quality is mandatory.
Materials and methods: Overall, one zirconia and four titanium dental implants were provided with different single crown materials: porcelain-fused-to-metal precious alloy (GP-T), porcelain-fused-to-metal non-precious alloy (CCT-T), porcelain-fused-to-zirconia (ZC-T) and monolithic zirconia (Z-T, Z-Z). Three-dimensional artefact volume was determined on a 3 Tesla MRI, applying two standard sequence types (SPACE and TSE). Two-way ANOVA and pair-wise post-hoc Turkey test were performed for comparison of artefact size.
Results: Fewest MR artefacts were observed with zirconia implant combined with monolithic zirconia crown. A titanium implant combined with a single crown framework out of the non-precious alloys was unfavourable in terms of artefact volume. Smaller and comparable artefact volumes were noted for titanium implants with the remaining three crown materials (GP-T, ZC-T and Z-T).
Conclusions: Material composition of dental implants provided with single crowns has a profound impact on artefact volume. In comparison with crowns containing cobalt, chromium and tungsten, the MRI artefacts are reduced in precious alloy- and zirconia-based crowns. Further studies are needed to assess whether residual artefacts allow sufficient diagnostic imaging with these crowns. Conflict-of-interest statement: The authors have nothing to disclose.