Purpose: This investigation aimed to develop a scanner quantification performance methodology and compare multiple metrics between two scanners under different imaging conditions. Most PET scanners are designed to work over a wide dynamic range of patient imaging conditions. Clinical constraints, however, often impact the realization of the entitlement performance for a particular scanner design. Using less injected dose and imaging for a shorter time are often key considerations, all while maintaining "acceptable" image quality and quantitative capability.
Methods: A dual phantom measurement including resolution inserts was used to measure the effects of in-plane (x, y) and axial (z) system resolution between two PET/CT systems with different block detector crystal dimensions. One of the scanners had significantly thinner slices. Several quantitative measures, including feature contrast recovery, max/min value, and feature profile accuracy were derived from the resulting data and compared between the two scanners and multiple phantoms and alignments.
Results: At the clinically relevant count levels used, the scanner with thinner slices had improved performance of approximately 2%, averaged over phantom alignments, measures, and reconstruction methods, for the head-sized phantom, mainly demonstrated with the rods aligned perpendicular to the scanner axis. That same scanner had a slightly decreased performance of -1% for the larger body-size phantom, mostly due to an apparent noise increase in the images. Most of the differences in the metrics between the two scanners were less than 10%.
Conclusions: Using the proposed scanner performance methodology, it was shown that smaller detector elements and a larger number of image voxels require higher count density in order to demonstrate improved image quality and quantitation. In a body imaging scenario under typical clinical conditions, the potential advantages of the design must overcome increases in noise due to lower count density.