Background: AeroForm is a new type of remote-controlled, needle-free, carbon dioxide-based expander involving a potentially faster method of tissue expansion. Results are presented here from the AirXpanders Patient Activated Controlled Tissue Expander pivotal trial comparing AeroForm to saline tissue expanders.
Methods: Women undergoing two-stage breast reconstruction were randomized at 17 U.S. sites in this U.S. Food and Drug Administration-approved investigational device exemption trial. Expansion in the investigational arm was performed by the patient in 10-cc increments up to 30 cc/day of carbon dioxide and in the control arm by the physician with periodic bolus injections of saline. Safety endpoints, expansion and reconstruction times, pain, and satisfaction were assessed.
Results: One hundred fifty women were treated: 98 with carbon dioxide expanders (n = 168) and 52 with saline expanders (n = 88). The treatment success rate (all breasts exchanged successfully excluding non-device-related failures) was 96.1 percent for carbon dioxide and 98.8 percent for saline. Median time to full expansion and completion of the second-stage operation was 21.0 and 108.5 days (carbon dioxide) versus 46.0 and 136.5 days (saline), respectively, with a similar rate of overall complications. Ease of use for the carbon dioxide expander was rated high by patients (98 percent) and physicians (90 percent).
Conclusions: The AirXpanders Patient Activated Controlled Tissue Expander trial results demonstrate that a carbon dioxide-based expander is an effective method of tissue expansion with a similar overall adverse event rate compared to saline expanders, and provides a more convenient and expedient expansion.
Clinical question/level of evidence: Therapeutic, I.