A technique for safer and more effective pulsed laser lithotripsy of urinary and biliary calculi was investigated in vitro. The technique involves enclosing the distal end of the laser delivery fiber in a "plasma shield." The plasma shield is a specially designed metal cap that serves to transfer the laser-induced mechanical impulse to the calculus while shielding surrounding tissue from direct laser exposure and thermal radiation. The metal cap also offers the advantage of effectively blunting the sharp fiber tip and improving its visualization under fluoroscopy. Plasma shield lithotripsy using a 200 micron quartz fiber inserted into a section of a modified 0.034 in. diameter stainless steel guide wire was tested in vitro on a variety of calculi and compared with results obtained using a 200 micron laser fiber applied directly. Calculi tested included cystine, struvite and calcium oxalate dihydrate urinary stones and pigmented cholesterol gallstones. The laser source was a flashlamp-pumped dye laser producing pulses of 1.2 microsecond duration and operated at a wavelength of 504 nm and pulse repetition frequency of 5 Hz. The results show that plasma shield lasertripsy is as effective as direct lasertripsy for fragmenting gallstones, struvite and calcium oxalate dihydrate calculi, is potentially safer, and can fragment cystine calculi which the pulsed dye laser applied directly cannot.