A novel stent inflation protocol improves long-term outcomes compared with rapid inflation/deflation deployment method

Catheter Cardiovasc Interv. 2017 Aug 1;90(2):233-240. doi: 10.1002/ccd.26930. Epub 2017 Jan 23.

Abstract

Background: High-pressure inflation for coronary stent deployment is universally performed. However, the duration of inflation is variable and does not take into account differences in lesion compliance. We developed a standardized "pressure optimization protocol" (POP) using inflation pressure stability rather than an arbitrary inflation time or angiographic balloon appearance for stent deployment. Whether this approach improves long-term outcomes is unknown.

Methods and results: 792 patients who underwent PCI using either rapid inflation/deflation (n = 376) or POP (n = 416) between January 2009 and March 2014 were included. Exclusion criteria included PCI for acute myocardial infarction, in-stent restenosis, chronic total occlusion, left main, and saphenous vein graft lesions. Primary endpoint was target vessel failure [TVF = combined end point of target vessel revascularization (TVR), myocardial infarction, and cardiac death]. Outcomes were analyzed in the entire cohort and in a propensity analysis. Stent implantation using POP with a median follow-up of 1317 days was associated with lower TVF compared with rapid inflation/deflation (10.1 vs. 17.8%, P < 0.0001). This difference was driven by a decrease in TVR (7 vs. 10.6%, P = 0.0016) and cardiac death (2.9 vs. 5.8%, P = 0.017) while there was no difference in myocardial infarction (1 vs. 1.9%, P = 0.19). In the Cox regression model, deployment using POP was the only independent predictor of reduced TVF (HR 0.43; 0.29-0.64; P < 0.0001). In the propensity analysis (330 patients per group) TVF remained lower with POP vs. rapid inflation/deflation (10 vs. 18%, P < 0.0001).

Conclusions: Stent deployment using POP led to reduced TVF compared to rapid I/D. These results recommend this method to improve long-term outcomes. © 2017 Wiley Periodicals, Inc.

Keywords: percutaneous coronary intervention; prolonged stent inflation; stent expansion; target vessel failure.

Publication types

  • Comparative Study

MeSH terms

  • Aged
  • Angioplasty, Balloon, Coronary / adverse effects
  • Angioplasty, Balloon, Coronary / instrumentation*
  • Angioplasty, Balloon, Coronary / methods*
  • Angioplasty, Balloon, Coronary / mortality
  • Chi-Square Distribution
  • Coronary Artery Disease / diagnostic imaging
  • Coronary Artery Disease / mortality
  • Coronary Artery Disease / therapy*
  • Female
  • Humans
  • Kaplan-Meier Estimate
  • Logistic Models
  • Male
  • Middle Aged
  • Myocardial Infarction / etiology
  • Propensity Score
  • Proportional Hazards Models
  • Retrospective Studies
  • Risk Factors
  • Stents*
  • Time Factors
  • Treatment Outcome