Label-Free Detection of Gliadin in Food by Quartz Crystal Microbalance-Based Immunosensor

J Agric Food Chem. 2017 Feb 15;65(6):1281-1289. doi: 10.1021/acs.jafc.6b04830. Epub 2017 Feb 6.

Abstract

Gluten is a protein composite found in wheat and related grains including barley, rye, oat, and all their species and hybrids. Gluten matrix is a biomolecular network of gliadins and glutenins that contribute to the texture of pastries, breads, and pasta. Gliadins are mainly responsible for celiac disease, one of the most widespread food-related pathologies in Western world. In view of the importance of gliadin proteins, by combining the quartz crystal microbalance technology, a cheap and robust piezoelectric transducer, with the so-called photonic immobilization technique, an effective surface functionalization method that provides spatially oriented antibodies on gold substrates, we realized a sensitive and reliable biosensor for quantifying these analytes extracted from real samples in a very short time. The resulting immunosensor has a limit of detection of about 4 ppm and, more remarkably, shows excellent sensitivity in the range 7.5-15 ppm. This feature makes our device reliable and effective for practical applications since it is able to keep low the influence of false positives.

Keywords: antibodies; gliadin; immunosensor; photonic immobilization technique; quartz-crystal microbalance; surface functionalization.

MeSH terms

  • Biosensing Techniques / instrumentation
  • Biosensing Techniques / methods
  • Blotting, Western
  • Chromatography, High Pressure Liquid / methods
  • Electrophoresis, Polyacrylamide Gel
  • Enzyme-Linked Immunosorbent Assay / methods
  • Food Analysis / instrumentation
  • Food Analysis / methods*
  • Gliadin / analysis*
  • Limit of Detection
  • Quartz Crystal Microbalance Techniques / instrumentation
  • Quartz Crystal Microbalance Techniques / methods*
  • Sensitivity and Specificity
  • Ultraviolet Rays

Substances

  • Gliadin