Cancer is a global health issue that disproportionately kills based on stage of disease, cellular pathology, and genetics, to name a few. Another variable to consider in this ongoing fight is treatment machine complexity that leads to elevated development and purchasing cost, leading to a reduced use. Reducing the complexity (in hopes of lowering costs) would benefit underdeveloped, low- and middle-income countries by introducing newer treatment technology, as their currently accepted standards do not meet standards of more advanced, developed countries. In this study, unilateral head and neck (H&N), and prostate cases using volumetric modulated arc therapy (VMAT) were tested with multiple segment widths of 5, 10, 15, and 20mm to create treatable plans. Pinnacle 9.10v was used for planning purposes. A total of 12 cases were planned with varying multileaf collimator (MLC) widths. Treatment plans were evaluated retrospectively. Results show that altering the MLC widths from 5 through 20mm produces both comparable and treatable plans up to 99% and 98% target coverage for H&N and prostate, respectively, albeit clinically significant hot spots were shown to increase with increasing segment width. Furthermore, the results show that increasing widths can produce comparable treatment plans as measured against our current Food and Drug Administration (FDA)-approved treatment devices-leading to an increase in treatment efficacy in economically underdeveloped countries.
Keywords: Cost; MLC (multileaf collimator) Segment Width; Underdeveloped countries; VMAT (volumetric modulated arc therapy).
Copyright © 2017 American Association of Medical Dosimetrists. Published by Elsevier Inc. All rights reserved.