A phantom study: Should 124 I-mIBG PET/CT replace 123 I-mIBG SPECT/CT?

Med Phys. 2017 May;44(5):1624-1631. doi: 10.1002/mp.12202. Epub 2017 Apr 17.

Abstract

Purpose: The isotope 123 I is commonly labeled with meta-iodobenzylguanidine (mIBG) for imaging of neuroendocrine tumors, such as pheochromocytomas and neuroblastomas. 123 I-mIBG SPECT/CT imaging is performed for staging, follow-up and selection of patients for treatment with 131 I mIBG. As an alternative to 123 I, 124 I-mIBG PET/CT may be used, potentially taking advantage of the superior PET image quality. The purpose of this study was to investigate whether 124 I PET/CT improves image quality as compared with 123 I SPECT/CT for equal patient effective radiation dose (in mSv).

Methods: Phantom measurements were performed using the NEMA-2007 image quality phantom. SPECT and PET reconstruction settings were used with and without time-of-flight (TOF) and point-spread-function (PSF) modeling. As a measure of image quality, the contrast-to-noise ratio (CNR) was calculated. The ratio of the 123 I to 124 I activity concentration was determined at which the contrast-to-noise ratio was equal for both modalities. This metric was defined as the contrast equivalent activity ratio (CEAR).

Results: CEARs of 47.7, 25.6, 23.1, 14.6, 10.0, and 9.1 were obtained for a TOF and PSF modeled 124 I reconstruction method and an attenuation and scatter-corrected 123 I reconstruction method for sphere sizes of 10 to 37 mm, respectively. As the effective radiation dose of 124 I-mIBG is higher than of 123 I-mIBG (in mSv/MBq), an equal effective dose corresponds to a CEAR of 5 to 10. Therefore, CEARs higher than 5 to 10 indicate that 124 I PET/CT outperforms 123 I SPECT/CT in the sense of image quality for equal patient effective radiation dose.

Conclusion: The CEAR is much larger than a factor of 5 to 10 (needed for equal patient effective radiation dose) for most of the reconstruction methods and sphere sizes. Therefore, 124 I-mIBG PET/CT is expected to improve image quality and/or may be used to reduce effective patient dose as compared with 123 I-mIBG SPECT/CT.

Keywords: 123I SPECT/CT; 124I PET/CT; CNR; image quality; mIBG; phantom.

MeSH terms

  • 3-Iodobenzylguanidine / pharmacokinetics*
  • Humans
  • Neuroendocrine Tumors / diagnostic imaging
  • Phantoms, Imaging
  • Positron Emission Tomography Computed Tomography*
  • Positron-Emission Tomography*
  • Tomography, Emission-Computed, Single-Photon
  • Tomography, X-Ray Computed

Substances

  • 3-Iodobenzylguanidine