Introduction: Factor XIII (FXIII) deficiency is a rare hemorrhagic disorder whose early diagnosis is crucial for appropriate treatment and prophylactic supplementation in cases of severe deficiency. International guidelines recommend a quantitative FXIII activity assay as first-line screening test. FXIII antigen measurement may be performed to establish the subtype of FXIII deficiency (FXIIID) when activity is decreased.
Methods: The aim of this multicenter study was to evaluate the analytical and diagnostic levels of performance of a new latex immunoassay, K-Assay® FXIII reagent from Stago, for first-line measurement of FXIII antigen. Results were compared to those obtained with the Berichrom® FXIII chromogenic assay for measurement of FXIII activity. Of the 147 patient plasma samples, 138 were selected for analysis.
Results: The accuracy was very good, with intercenter reproducibility close to 7%. Five groups were defined on FXIII activity level (<5% (n = 5), 5%-30% (n = 23), 30%-60% (n = 17), 60%-120% (n = 69), above 120% (n = 24)), without statistical differences between activity and antigen levels (P value >0.05). Correlation of the K-Assay® with the Berichrom® FXIII activity results was excellent (r = 0.919). Good agreement was established by the Bland and Altman method, with a bias of +9.4% on all samples, and of -1.4% for FXIII levels lower than 30%. One patient with afibrinogenemia showed low levels of Berichrom® FXIII activity but normal antigen level and clot solubility as expected.
Conclusions: The measurement of FXIII antigen using the K-Assay® is a reliable first-line tool for detection of FXIII deficiency when an activity assay is not available.
Keywords: FXIII activity; FXIII antigen; FXIII deficiency.
© 2017 John Wiley & Sons Ltd.