A newly acquired nanoDot In-Light system was compared with TLD-100 dosimeters to confirm the treatment dose in the multiple cases: an electron eye treatment, H&N IMRT and VMAT validation for small targets. Eye tumour treatment with 9 MeV electrons A dose of 1.8 Gy per fraction was prescribed to the 85% isodose. The average dose measured by three TLDs and three Dots was 1.90 and 1.97 Gy. Both detectors overestimated dose, by 2.9% and 6.7% respectively. H&N IMRT treatment of skin cancer with 6 MV photons Dose per fraction is 2.5 Gy. The average doses measured by two TLDs and two Dots were 2.48 and 2.56 Gy, which represent errors of -0.8% and 2.2%, respectively. VMAT validation for small targets using an Agarose phantom, dose 15 Gy A single-tumour brain treatment was delivered using two coplanar arcs to an Agarise phantom containing a large plastic insert holding 3 nanoDots and 4 TLDs. The difference between the average Pinnacle dose and the average dose of the corresponding detectors was -0.6% for Dots and -1.7% for TLDs. A two-tumour brain treatment was delivered using three non-coplanar arcs. Small and large plastic inserts separated by 5 cm were used to validate the dose. The difference between the average Pinnacle dose and the average dose of the corresponding detectors was the following; small phantom 0.7% for Dots and 0.3% for TLDs, large phantom-1.9% for Dots and -0.6% for TLDs. In conclusion, nanoDot detectors are suitable for in-vivo dosimetry with photon and electron beams.
Keywords: Brain; Cancer; Dosimetry; Electron beams; Intensity modulated radiation therapy; Nanodots; Particle beam detectors; Photons; Thermoluminescent dosimeters.