[Technique of complex mammary irradiation: Mono-isocentric 3D conformational radiotherapy and helical tomotherapy]

Cancer Radiother. 2017 Dec;21(8):741-748. doi: 10.1016/j.canrad.2017.03.005. Epub 2017 Jul 8.
[Article in French]

Abstract

Purpose: To evaluate the dosimetric contribution of helical tomotherapy for breast cancers compared with conformal radiotherapy in mono-isocentric technique.

Patients and method: For 23 patients, the dosimetric results in mono-isocentric 3D conformational radiotherapy did not satisfy the constraints either of target volumes nor organs at risk. A prospective dosimetric comparison between mono-isocentric 3D conformational radiotherapy and helical tomotherapy was therefore carried out.

Results: The use of helical tomotherapy showed a benefit in these 23 patients, with either an improvement in the conformity index or homogeneity, but with an increase in low doses. Of the 23 patients, two had pectus excavatum, five had past thoracic irradiation and two required bilateral irradiation. The other 14 patients had a combination of morphology and/or indication of lymph node irradiation. For these patients, helical tomotherapy was therefore preferred to mono-isocentric 3D conformational radiotherapy.

Conclusions: Tomotherapy appears to provide better homogeneity and tumour coverage. This technique of irradiation may be justified in the case of morphological situations such as pectus exavatum and in complex clinical situations. In other cases, conformal radiotherapy in mono-isocentric technique remains to be favoured.

Keywords: Breast cancer; Cancer du sein; Mono-isocentric technic; Technique monoisocentrique; Tomotherapy; Tomothérapie.

Publication types

  • Comparative Study

MeSH terms

  • Adult
  • Aged
  • Breast Neoplasms / radiotherapy*
  • Female
  • Humans
  • Middle Aged
  • Prospective Studies
  • Radiotherapy Dosage
  • Radiotherapy, Conformal*
  • Radiotherapy, Intensity-Modulated