Effect of Emergency Department Mattress Compressibility on Chest Compression Depth Using a Standardized Cardiopulmonary Resuscitation Board, a Slider Transfer Board, and a Flat Spine Board: A Simulation-Based Study

Simul Healthc. 2017 Dec;12(6):364-369. doi: 10.1097/SIH.0000000000000245.

Abstract

Introduction: Cardiopulmonary resuscitation (CPR) performed on a mattress decreases effective chest compression depth. Using a CPR board partially attenuates mattress compressibility. We aimed to determine the effect of a CPR board, a slider transfer board, a CPR board with a slider transfer board, and a flat spine board on chest compression depth with a mannequin placed on an emergency department mattress.

Methods: The study used a cross-over study design. The CPR-certified healthcare providers performed 2 minutes of compressions on a mannequin in five conditions, an emergency department mattress with: (a) no hard surface, (b) a CPR board, (c) a slider transfer board, (d) a CPR board and slider transfer board, and (e) a flat spine board. Compression depths were measured from two sources for each condition: (a) an internal device measuring sternum-to-spine compression and (b) an external device measuring sternum-to-spine compression plus mattress compression. The difference of the two measures (ie, depleted compression depth) was summarized and compared between conditions.

Results: A total of 10,203 individual compressions from 10 participants were analyzed. The mean depleted compression depths (percentage depletion) secondary to mattress effect were the following: 23.6 mm (29.7%) on a mattress only, 13.7 mm (19.5%) on a CPR board, 16.9 mm (23.1%) on a slider transfer board, 11.9 mm (17.3%) on a slider transfer board plus backboard, and 10.3 mm (15.4%) on a flat spine board. The differences in percentage depletion across conditions were statistically significant.

Conclusion: Cardiopulmonary resuscitation providers should use a CPR board and slider transfer board or a flat spine board alone because these conditions are associated with the smallest amount of mattress compressibility.

Publication types

  • Clinical Trial

MeSH terms

  • Beds*
  • Cardiopulmonary Resuscitation / instrumentation*
  • Cross-Over Studies
  • Emergency Service, Hospital / organization & administration*
  • Equipment Design