Acute and midterm results following perventricular device closure of muscular ventricular septal defects: A multicenter PICES investigation

Catheter Cardiovasc Interv. 2017 Aug 1;90(2):281-289. doi: 10.1002/ccd.27121. Epub 2017 May 8.

Abstract

Objectives: To describe acute and mid-term results of hybrid perventricular device closure of muscular ventricular septal defects (mVSDs).

Background: Perventricular device closure of mVSDs can mitigate technical limitations of percutaneous closure and need for cardiopulmonary bypass or ventriculotomy with a surgical approach.

Methods: This is a multicenter retrospective cohort study of patients undergoing hybrid perventricular mVSD device closure from 1/2004 to 1/2014. Procedural details, adverse events, outcomes, and follow-up data were collected. Patients were divided into two groups: (1) simple (mVSD closure alone) and (2) complex (mVSD closure with concomitant cardiac surgery).

Results: Forty-seven patients (60% female) underwent perventricular mVSD device closure at a median age of 5.2 months (IQR 1.8-8.9) and weight of 5.1 kg (IQR 4.0-6.9). Procedural success was 91% [100% (n = 22) simple and 84% (n = 21/25) complex]. Adverse events occurred in 19% (9/47) [9% (2/22) simple and 28% (7/25) complex]. Hospital length of stay (LOS) was shorter in the simple vs. complex group (4 vs. 14 days, P < 0.01). At mid-term follow-up of 19.2 months (IQR 2.3-43) 90% of pts had complete mVSD closure; none developed late heart block, increased atrioventricular (AV) valve insufficiency or ventricular dysfunction.

Conclusions: Perventricular device closure of simple mVSD was associated with a high rate of procedural success, few adverse events, and short hospital LOS. Procedural adverse events were associated with the presence of concomitant complex surgery. Residual mVSD, AV valve insufficiency, or ventricular dysfunction were uncommon at mid-term follow-up. © 2017 Wiley Periodicals, Inc.

Keywords: VSD device closure; hybrid; midterm outcomes; muscular ventricular septal defects; perventricular.

Publication types

  • Multicenter Study

MeSH terms

  • Cardiac Surgical Procedures* / adverse effects
  • Child
  • Child, Preschool
  • Echocardiography, Transesophageal
  • Female
  • Heart Septal Defects, Ventricular / diagnostic imaging
  • Heart Septal Defects, Ventricular / therapy*
  • Humans
  • Infant
  • Length of Stay
  • Male
  • Percutaneous Coronary Intervention / adverse effects
  • Percutaneous Coronary Intervention / instrumentation*
  • Prosthesis Design
  • Retrospective Studies
  • Risk Factors
  • Septal Occluder Device*
  • Time Factors
  • Treatment Outcome
  • United States