Purpose: To compare the detection rate of choroidal neovascularization (CNV) in treatment-naive neovascular age-related macular degeneration by swept source optical coherence tomography angiography (SS-OCTA, Topcon's DRI Triton) working at 1,050 nm wavelength versus fluorescence angiography.
Methods: Cross-sectional analysis of 156 eyes (107 neovascular age-related macular degeneration and 49 dry AMD) in 98 patients, previously diagnosed by multimodal imaging using fluorescein (FA) and indocyanine green angiography (Heidelberg's Spectralis) in a tertiary retina center, evaluated by SS-OCTA 4.5 mm × 4.5 mm and 6 mm × 6 mm macular cubes. Main outcome measures were sensitivity and specificity of SS-OCTA in AMD. Potential factors influencing CNV detection rate were analyzed.
Results: Swept source optical coherence tomography angiography detected CNV in 81 of 107 eyes, resulting in a sensitivity of 75.7%. In 49 eyes with dry AMD, no CNV could be identified (specificity 100%). A statistical significance was calculated for nondetection of treatment-naive CNV by SS-OCTA in pigment epithelial detachment over 400 μm (P = 0.0238).
Conclusion: Topcon's SS-OCTA was not able to detect all CNV lesions. Large pigment epithelial detachments were associated with signal loss. Fluorescence angiography still remains the gold standard, but the tested SS-OCTA device can be considered as a feasible additional diagnostic tool in AMD.