Introduction: Early diagnosis of HIV-1 infection and the prompt initiation of antiretroviral therapy are critical to achieving a reduction in the morbidity and mortality of infected infants. The Simple AMplification-Based Assay (SAMBA) HIV-1 Qual Whole Blood Test was developed specifically for early infant diagnosis and prevention of mother-to-child transmission programs implemented at the point-of-care in resource-limited settings.
Methods: We have evaluated the performance of this test run on the SAMBA I semiautomated platform with fresh whole blood specimens collected from 202 adults and 745 infants in Kenya, Uganda, and Zimbabwe. Results were compared with those obtained with the Roche COBAS AmpliPrep/COBAS TaqMan (CAP/CTM) HIV-1 assay as performed with fresh whole blood or dried blood spots of the same subjects, and discrepancies were resolved with alternative assays.
Results: The performance of the SAMBA and CAP/CTM assays evaluated at 5 laboratories in the 3 countries was similar for both adult and infant samples. The clinical sensitivity, specificity, positive predictive value, and negative predictive value for the SAMBA test were 100%, 99.2%, 98.7%, and 100%, respectively, with adult samples, and 98.5%, 99.8%, 99.7%, and 98.8%, respectively, with infant samples.
Discussion: Our data suggest that the SAMBA HIV-1 Qual Whole Blood Test would be effective for early diagnosis of HIV-1 infection in infants at point-of-care settings in sub-Saharan Africa.