Very Late Pathological Responses to Cobalt-Chromium Everolimus-Eluting, Stainless Steel Sirolimus-Eluting, and Cobalt-Chromium Bare Metal Stents in Humans

J Am Heart Assoc. 2017 Nov 17;6(11):e007244. doi: 10.1161/JAHA.117.007244.

Abstract

Background: The "very late" clinical outcomes for durable polymer drug-eluting stents and bare metal stents (BMSs) have been shown to be dissimilar in clinical studies. Conceptually, the long-term vascular compatibility of BMSs is still regarded to be superior to drug-eluting stents; however, no pathologic study to date has specifically addressed this issue. We evaluated the very late (≥1 year) pathologic responses to durable polymer drug-eluting stents (cobalt-chromium [CoCr] everolimus-eluting stents [EESs] and stainless steel sirolimus-eluting stents [SS-SESs]) versus BMSs (CoCr-BMSs).

Methods and results: From the CVPath stent registry, we studied a total of 119 lesions (40 CoCr-EESs, 44 SS-SESs, 35 CoCr-BMSs) from 92 autopsy cases with a duration ranging from 1 to 5 years. Sections of stented coronary segments were pathologically analyzed. Inflammation score and the percentage of struts with giant cells were lowest in CoCr-EESs (median inflammation score: 0.6; median percentage of struts with giant cells: 3.8%) followed by CoCr-BMSs (median inflammation score: 1.3 [P<0.01]; median percentage of struts with giant cells: 8.9% [P=0.02]) and SS-SESs (median inflammation score: 1.7 [P<0.01]; median percentage of struts with giant cells: 15.3% [P<0.01]). Polymer delamination was observed exclusively in SS-SESs and was associated with increased inflammatory and giant cell reactions. The prevalence of neoatherosclerosis with CoCr-EESs (50%) was significantly less than with SS-SESs (77%, P=0.02) but significantly greater than with CoCr-BMSs (20%, P<0.01).

Conclusions: CoCr-EESs, SS-SESs, and BMSs each demonstrated distinct vascular responses. CoCr-EESs demonstrated the least inflammation, near-equivalent healing to BMSs, and lower neointimal formation. These results challenge the belief that BMSs have superior biocompatibility compared with some polymeric coated drug-eluting stents and may have implications for future stent design.

Keywords: bare metal stent; drug‐eluting stent; inflammation; neoatherosclerosis; pathology.

MeSH terms

  • Chromium Alloys*
  • Coronary Artery Disease / diagnostic imaging
  • Coronary Artery Disease / surgery*
  • Coronary Vessels / pathology*
  • Drug-Eluting Stents*
  • Female
  • Follow-Up Studies
  • Humans
  • Male
  • Middle Aged
  • Registries*
  • Retrospective Studies
  • Time Factors
  • Tomography, Optical Coherence
  • Treatment Outcome

Substances

  • Chromium Alloys