Clinical implementation of contrast-enhanced four-dimensional dual-energy computed tomography for target delineation of pancreatic cancer

Radiother Oncol. 2018 Oct;129(1):105-111. doi: 10.1016/j.radonc.2018.01.012. Epub 2018 Feb 6.

Abstract

Background and purpose: The accurate delineation of pancreatic tumor with respiratory motion is challenging. This study demonstrates the application of contrast-enhanced four-dimensional dual-energy computed tomography (CE-4D-DECT) for tumor delineation and assesses the objective and subjective image quality.

Material and methods: Twelve patients underwent CE-4D-DECT, and quantitative spectral analysis was performed on the resulting virtual monochromatic images (VMI) to determine the optimal VMI (O-VMI) with the highest contrast-to-noise ratio (CNR). The objective value of the CNR between pancreatic parenchyma and tumor, and the subjective measurement with five-point scale were compared between O-VMI, standard VMI (S-VMI, 77 keV) and single energy CT (SECT, 120 kVp).

Results: The CNR was the highest in the VMI at 60 keV, and the corresponding CNR in the O-VMI (3.4) was significantly higher (p < 0.05) than that in the S-VMI (2.4) and the SECT (2.7). The overall mean subjective measurements among 4 radiation oncologists were higher for the O-VMI over the S-VMI and SECT with respect to overall image quality (4.0, 3.3 and 3.7, respectively), tumor enhancement (3.4, 2.6 and 3.2, respectively), and vessel delineation (4.2, 3.6 and 4.2, respectively).

Conclusions: The O-VMI derived from the CE-4D-DECT demonstrated its superiority over the S-VMI and SECT in depicting pancreatic tumor.

Keywords: 4DCT; Contrast enhanced; Dual energy; Pancreatic cancer.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Aged
  • Aged, 80 and over
  • Female
  • Four-Dimensional Computed Tomography / methods*
  • Humans
  • Male
  • Middle Aged
  • Pancreatic Neoplasms / diagnostic imaging*
  • Radiographic Image Interpretation, Computer-Assisted / methods
  • Radiography, Dual-Energy Scanned Projection / methods*
  • Retrospective Studies
  • Signal-To-Noise Ratio