Background: IQ-SPECT has been shown to significantly reduce acquisition time and administered dose while preserving image quality in myocardial perfusion imaging. Whether IQ-SPECT provides accurate left ventricular ejection fractions (LVEF) with gated blood pool SPECT (GBPS) remains unknown.
Methods: Sixty patients underwent IQ-SPECT GBPS and planar imaging. Among those patients, 11 underwent both cMRI and GBPS. GBPS LVEF, LVEDV, and LVESV were calculated using 2 validated software; QBS (Cedars-Sinai Medical Center, Los Angeles, USA) and MHI (Montreal Heart Institute, Montreal, Canada). LVEF, LVEDV, and LVESV obtained with the different modalities were compared.
Results: Average planar LVEF was 48 ± 11% (mean ± SD), average LVEDV was 177 ± 59 mL (range 63 to 342 mL), and average LVESV was 96 ± 46 mL (range 16 to 234 mL). GBPS LVEF and their correlation coefficient with planar LVEF were 40 ± 12% (r = 0.70) and 44 ± 12% (r = 0.83) with QBS and MHI, respectively. Correlation coefficient between cMRI and planar LVEF was 0.65 and were 0.69 and 0.52 between cMRI and GBPS using QBS and MHI, respectively.
Conclusions: LVEF calculated with GBPS using IQ-SPECT correlates with planar measurements. Correlation is best using the MHI method and variation is independent of LVEDV.
Keywords: Collimation; Gated blood pool imaging; Left ventricular function; Radionuclide angiography.