Comparison of Enzymatic Assay for HBA1C Measurement (Abbott Architect) With Capillary Electrophoresis (Sebia Minicap Flex Piercing Analyser)

Lab Med. 2018 Jul 5;49(3):231-238. doi: 10.1093/labmed/lmx090.

Abstract

Objective: To compare the analytical performances of the enzymatic method (EM) and capillary electrophoresis (CE) for hemoglobin A1c (HbA1c) measurement.

Methods: Imprecision, carryover, stability, linearity, method comparison, and interferences were evaluated for HbA1c via EM (Abbott Laboratories, Inc) and CE (Sebia).

Results: Both methods have shown overall within-laboratory imprecision of less than 3% for International Federation of Clinical Chemistry and Laboratory Medicine (IFCC) units (<2% National Glycohemoglobin Standardization Program [NGSP] units). Carryover effects were within acceptable criteria. The linearity of both methods has proven to be excellent (R2 = 0.999). Significant proportional and constant difference were found for EM, compared with CE, but were not clinically relevant (<5 mmol/mol; NGSP <0.5%). At the clinically relevant HbA1c concentration, stability observed with both methods was acceptable (bias, <3%). Triglyceride levels of 8.11 mmol per L or greater showed to interfere with EM and fetal hemoglobin (HbF) of 10.6% or greater with CE.

Conclusion: The enzymatic method proved to be comparable to the CE method in analytical performances; however, certain interferences can influence the measurements of each method.

Publication types

  • Comparative Study

MeSH terms

  • Electrophoresis, Capillary / methods
  • Electrophoresis, Capillary / standards*
  • Enzyme Assays / methods*
  • Enzyme Assays / standards*
  • Glycated Hemoglobin / analysis*
  • Humans
  • Linear Models
  • Reproducibility of Results

Substances

  • Glycated Hemoglobin A