Chronic Apical and Nonapical Right Ventricular Pacing in Patients with High-Grade Atrioventricular Block: Results of the Right Pace Study

Biomed Res Int. 2018 May 22:2018:1404659. doi: 10.1155/2018/1404659. eCollection 2018.

Abstract

Objective: The aim of the study was to compare the two approaches to chronic right ventricular pacing currently adopted in clinical practice: right ventricular apical (RVA) and non-RVA pacing.

Background: Chronic RVA pacing is associated with an increased risk of atrial fibrillation, morbidity, and even mortality. Non-RVA pacing may yield more physiologic ventricular activation and provide potential long-term benefits and has recently been adopted as standard procedure at many implanting centers.

Methods: The Right Pace study was a multicenter, prospective, single-blind, nonrandomized trial involving 437 patients indicated for dual-chamber pacemaker implantation with a high percentage of RV pacing.

Results: RV lead-tip target location was the apex or the interventricular septum. RVA (274) and non-RVA patients (163) did not differ in baseline characteristics. During a median follow-up of 19 months (25th-75th percentiles, 13-25), 17 patients died. The rates of the primary outcome of death due to any cause or hospitalization for heart failure were comparable between the groups (log-rank test, p = 0.609), as were the rates of the composite of death due to any cause, hospitalization for heart failure, or an increase in left ventricular end-systolic volume ≥ 15% as compared with the baseline evaluation (secondary outcome, p = 0.703). After central adjudication of X-rays, comparison between adjudicated RVA (239 patients) and non-RVA (170 patients) confirmed the absence of difference in the rates of primary (p = 0.402) and secondary (p = 0.941) outcome.

Conclusions: In patients with indications for dual-chamber pacemaker who require a high percentage of ventricular stimulation, RVA or non-RVA pacing resulted in comparable outcomes. This study is registered with ClinicalTrials.gov (identifier: NCT01647490).

Publication types

  • Clinical Trial
  • Multicenter Study

MeSH terms

  • Aged
  • Aged, 80 and over
  • Atrioventricular Block*
  • Cardiac Pacing, Artificial*
  • Female
  • Heart Ventricles
  • Humans
  • Male
  • Prospective Studies
  • Quality of Life
  • Reproducibility of Results
  • Single-Blind Method
  • Treatment Outcome

Associated data

  • ClinicalTrials.gov/NCT01647490