Multicenter Evaluation of Prosthesis Oversizing of the SAPIEN 3 Transcatheter Heart Valve. Impact on Device Failure and New Pacemaker Implantations
Rev Esp Cardiol (Engl Ed). 2019 Aug;72(8):641-648.
doi: 10.1016/j.rec.2018.06.005.
Epub 2018 Jul 17.
[Article in
English,
Spanish]
Authors
Costanza Pellegrini
1
, Won-Keun Kim
2
, Andreas Holzamer
3
, Thomas Walther
4
, N Patrick Mayr
5
, Jonathan Michel
1
, Tobias Rheude
1
, Julio Nuñez
1
, Albert M Kasel
1
, Teresa Trenkwalder
1
, Bernhard M Kaess
1
, Michael Joner
6
, Adnan Kastrati
6
, Heribert Schunkert
6
, Michael Hilker
3
, Helge Möllmann
7
, Christian Hengstenberg
8
, Oliver Husser
1
Affiliations
- 1 Klinik für Herz- und Kreislauferkrankungen, Deutsches Herzzentrum München, Technical University Munich, Munich, Germany.
- 2 Department of Cardiology, Kerckhoff Heart and Lung Center, Bad Nauheim, Germany; Department of Cardiovascular Surgery, Kerckhoff Heart and Lung Center, Bad Nauheim, Germany.
- 3 Klinik für Herz-, Thorax-, und herznahe Gefäßchirurgie, University of Regensburg Medical Center, Regensburg, Germany.
- 4 Department of Cardiovascular Surgery, Kerckhoff Heart and Lung Center, Bad Nauheim, Germany.
- 5 Institut für Anästhesiologie, Deutsches Herzzentrum München, Technical University Munich, Munich, Germany.
- 6 Klinik für Herz- und Kreislauferkrankungen, Deutsches Herzzentrum München, Technical University Munich, Munich, Germany; Deutsches Zentrum für Herz- und Kreislauf-Forschung (DZHK) e.V. (German Center for Cardiovascular Research), Partner Site Munich Heart Alliance, Munich, Germany.
- 7 Department of Cardiology, Kerckhoff Heart and Lung Center, Bad Nauheim, Germany.
- 8 Klinik für Herz- und Kreislauferkrankungen, Deutsches Herzzentrum München, Technical University Munich, Munich, Germany; Deutsches Zentrum für Herz- und Kreislauf-Forschung (DZHK) e.V. (German Center for Cardiovascular Research), Partner Site Munich Heart Alliance, Munich, Germany; Division of Cardiology, Department of Internal Medicine II, Medical University of Vienna, Vienna, Austria. Electronic address: [email protected].
Abstract
Introduction and objectives:
A certain degree of prosthesis oversizing (OS) is recommended for the SAPIEN 3 to achieve device success. However, an increase in OS may increase permanent pacemaker implantation (PPI) rates. We therefore investigated the influence of OS on device failure and PPI.
Methods:
A total of 804 patients were treated with SAPIEN 3 at 3 centers. Multislice computed tomography-derived OS was calculated and analyzed both as a continuous variable and categorized in 5% increments with -4% to 0% as reference.
Results:
Device failure occurred in 8.8% of patients. Median OS was lower in patients with device failure vs those with device success (+4% vs +8%; P=.038). A nonlinear risk pattern was shown for OS with a significantly reduced device failure rate within 4% to +22% of OS. There was no case of paravalvular leakage II+ between +10% to +20% of OS. The overall PPI rate was 16.2% and the median OS was significantly larger in patients with PPI (PPI: +9% vs no PPI: +7%; P = .025), while implantation depth did not vary in patients with vs without PPI (6.9±1.7 mm vs 6.6±1.9 mm; P=.101). The risk of PPI increased with increasing OS and was highest in the 2 highest categories.
Conclusions:
An increase in OS reduces the risk for device failure but increases the risk for PPI. There was no ideal range of OS to minimize both device failure and PPI.
Keywords:
Device failure; Dimensionamiento de la prótesis; Fallo de dispositivo; Implante percutáneo de válvula aórtica; Marcapasos; Pacemaker; Sizing; Transcatheter aortic valve implantation.
Copyright © 2018 Sociedad Española de Cardiología. Published by Elsevier España, S.L.U. All rights reserved.
MeSH terms
-
Aged, 80 and over
-
Aortic Valve / diagnostic imaging
-
Aortic Valve / surgery*
-
Aortic Valve Stenosis / diagnosis
-
Aortic Valve Stenosis / surgery*
-
Female
-
Follow-Up Studies
-
Heart Valve Prosthesis / adverse effects*
-
Humans
-
Male
-
Multidetector Computed Tomography
-
Pacemaker, Artificial*
-
Postoperative Complications / etiology
-
Postoperative Complications / therapy*
-
Prosthesis Failure
-
Retrospective Studies
-
Transcatheter Aortic Valve Replacement / adverse effects*
-
Treatment Outcome