Context.—: KRAS Mutation Test v2 is used for the qualitative detection and identification of 28 mutations in exons 2, 3, and 4 of the human KRAS gene.
Objective.—: To verify the performance of KRAS Mutation Test v2 and to evaluate its accuracy by correlation with a next-generation sequencing method on Illumina MiSeq.
Design.—: In this study, we used formalin-fixed, paraffin-embedded tissue and plasma specimens from non-small cell lung cancer, colorectal cancer, and pancreatic cancer patients. Results of specificity, precision, analytical sensitivity, and accuracy as compared with a MiSeq method are reported.
Results.—: The KRAS Mutation Test v2 demonstrated exquisite sensitivity and specificity and broad coverage of KRAS mutations. Precision was 100% (108 of 108) across all samples, operators, and instruments for formalin-fixed, paraffin-embedded tissue and 99.8% (615 of 616) for plasma. Analytical sensitivity was high with detection of 1% mutant sequence in formalin-fixed, paraffin-embedded tissue samples and as low as 25 mutant sequence copies/mL for plasma samples. The test also showed high overall concordance for formalin-fixed, paraffin-embedded tumor tissue as well as for plasma specimens when compared with MiSeq sequencing results.
Conclusions.—: The KRAS Mutation Test v2 is a highly robust, reproducible, and sensitive test for the qualitative detection of 28 mutations in exons 2, 3, and 4 of the KRAS gene in both solid (tissue) and liquid (plasma) biopsies from colorectal cancer, non-small cell lung cancer, and pancreatic cancer, and is a convenient option for KRAS mutation testing.