Background: Multiecho gradient-echo Cartesian MRI characterizes placental oxygenation by quantifying . Previous research was performed at 1.5T using breath-held 2D imaging during later gestational age (GA).
Purpose: To evaluate the accuracy and repeatability of a free-breathing (FB) 3D multiecho gradient-echo stack-of-radial technique (radial) for placental mapping at 3T and report placental during early GA.
Study type: Prospective.
Population: Thirty subjects with normal pregnancies and three subjects with ischemic placental disease (IPD) were scanned twice: between 14-18 and 19-23 weeks GA.
Field strength: 3T.
Sequence: FB radial.
Assessment: Linear correlation (concordance coefficient, ρc ) and Bland-Altman analyses (mean difference, MD) were performed to evaluate radial mapping accuracy compared to Cartesian in a phantom. Radial mapping repeatability was characterized using the coefficient of repeatability (CR) between back-to-back scans. The mean and spatial coefficient of variation (CV) of was determined for all subjects, and separately for anterior and posterior placentas, at each GA range.
Statistical tests: ρc was tested for significance. Differences in mean and CV were tested using Wilcoxon Signed-Rank and Rank-Sum tests. P < 0.05 was considered significant. Z-scores for the IPD subjects were determined.
Results: FB radial demonstrated accurate (ρc ≥0.996; P < 0.001; |MD|<0.2s-1 ) and repeatable (CR<4s-1 ) mapping in a phantom, and repeatable (CR≤4.6s-1 ) mapping in normal subjects. At 3T, placental mean ± standard deviation was 12.9s-1 ± 2.7s-1 for 14-18 and 13.2s-1 ± 1.9s-1 for 19-23 weeks GA. The CV was significantly greater (P = 0.043) at 14-18 (0.63 ± 0.12) than 19-23 (0.58 ± 0.13) weeks GA. At 19-23 weeks, the CV was significantly lower (P < 0.001) for anterior (0.49 ± 0.08) than posterior (0.67 ± 0.11) placentas. One IPD subject had a lower mean than normal subjects at both GA ranges (Z<-2).
Data conclusion: FB radial provides accurate and repeatable 3D mapping for the entire placenta at 3T during early GA.
Level of evidence: 2 Technical Efficacy: Stage 1 J. Magn. Reson. Imaging 2019;49:291-303.
Keywords: mapping; 3D radial imaging; 3T; early gestation; free-breathing MRI; placenta MRI.