Modern Dual-Mobility Cups in Revision Total Hip Arthroplasty: A Systematic Review and Meta-Analysis

J Arthroplasty. 2018 Dec;33(12):3793-3800. doi: 10.1016/j.arth.2018.08.013. Epub 2018 Aug 18.

Abstract

Background: The purpose of this systematic review is to analyze the outcomes of dual-mobility (DM) cups in revision total hip arthroplasty (THA). Specifically, we evaluated the following: (1) all-cause and aseptic survivorship rates; (2) dislocation rates; (3) complications; and (4) clinical outcomes reported using validated health status measures.

Methods: A comprehensive literature search included studies that reported the following: (1) re-revision rates, (2) complications, and (3) clinical outcomes following DM use in revision THA. The following exclusion criteria were used: (1) studies that did not stratify their analysis between primary and revision THA, (2) studies that utilized off-label techniques, (3) review articles, (4) case studies, (5) basic science articles, (6) non-English language reports, and (6) reports on patients who underwent surgery before 2010, in order to reflect modern DM implants use and technology. A total of 9 studies were included in our final analysis.

Results: Aseptic and all-cause survivorship rates were 97.7% and 94.5%. Prevalence of dislocation was 2.2%, and 0.3% for intraprosthetic dislocation. Meta-analysis comparing DM to fixed-bearing prostheses demonstrated a significantly lower odds of dislocation in the DM cohort (odds ratio 0.24, P = .002). Complications occurred in 7.4% of revision THAs with DM cups, while infection rates totaled to 3.3% of cases. Studies comparing outcomes using Harris Hip Scores did not demonstrate a statistically significant difference in improved postoperative scores (P > .05).

Conclusion: DM cups have demonstrated excellent survivorship, low dislocation, and overall complication rates. Therefore, it can be considered a safe and effective option, particularly in the high-risk patients who undergo revision THA.

Keywords: dislocations; dual mobility; fixed bearing; meta-analysis; total hip arthroplasty.

Publication types

  • Meta-Analysis
  • Systematic Review

MeSH terms

  • Arthroplasty, Replacement, Hip / instrumentation*
  • Arthroplasty, Replacement, Hip / statistics & numerical data
  • Health Status
  • Hip Prosthesis / adverse effects
  • Hip Prosthesis / statistics & numerical data*
  • Humans
  • Joint Dislocations
  • Odds Ratio
  • Postoperative Complications / epidemiology*
  • Postoperative Complications / etiology
  • Postoperative Period
  • Prosthesis Design
  • Prosthesis Failure
  • Range of Motion, Articular
  • Reoperation / instrumentation*
  • Reoperation / statistics & numerical data