Excellent Glycemic Control Maintained by Open-Source Hybrid Closed-Loop AndroidAPS During and After Sustained Physical Activity

Diabetes Technol Ther. 2018 Nov;20(11):744-750. doi: 10.1089/dia.2018.0214. Epub 2018 Oct 4.

Abstract

Objective: Officially licensed hybrid closed-loop systems are not currently available worldwide; therefore, open-source systems have become increasingly popular. Our aim was to assess the safety, feasibility, and efficacy of an open-source hybrid closed-loop system (AndroidAPS) versus SmartGuard® technology for day-and-night glucose control in children under extreme sports conditions.

Research design and methods: Twenty-two children (16 girls, 6-15 years of age, median HbA1c 56 ± 9 mmol/mol) were enrolled in this pivotal winter sports camp study. The participants were divided into two groups using either the AndroidAPS or SmartGuard technology. Physical exertion was represented by all-day alpine skiing. The primary endpoints were mean glucose level, time below the threshold of 3.9 mmol/L, and time within the target range of 3.9 to 10 mmol/L.

Results: The children using the AndroidAPS had significantly lower mean glycemia levels (7.2 ± 2.7 vs. 7.7 ± 2.8 mmol/L; 129.6 ± 49 vs. 138.6 ± 50 mg/dL, P < 0.042) than the children using the SmartGuard. The proportion of time below the target (median 5.0% ± 2.5% vs. 3.0% ± 2.3%, P = 0.6) and in the target zone (63% ± 9.5% vs. 63% ± 18%, P = 0.5) did not significantly differ. The AndroidAPS group experienced more frequent malfunctions of the cannula set (median 0.8 ± 0.4 vs. 0.2 ± 0.4, P = 0.02), which could have affected the results. No significant difference was found in the amount of carbohydrates consumed for the prevention and treatment of hypoglycemia [median 40 ± 23 vs. 25 ± 29 g/(patient ·3 days)]. No episodes of severe hypoglycemia or other serious adverse events were noted.

Conclusions: This pilot study showed that the AndroidAPS system was a safe and feasible alternative to the SmartGuard Technology.

Keywords: AndroidAPS; Do-it-yourself systems.; Open-source hybrid closed loop.

Publication types

  • Evaluation Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adolescent
  • Blood Glucose / metabolism
  • Blood Glucose Self-Monitoring / instrumentation*
  • Blood Glucose Self-Monitoring / methods
  • Child
  • Diabetes Mellitus, Type 1 / blood
  • Diabetes Mellitus, Type 1 / drug therapy*
  • Diabetes Mellitus, Type 1 / physiopathology
  • Exercise / physiology
  • Feasibility Studies
  • Female
  • Glycated Hemoglobin / metabolism
  • Humans
  • Hypoglycemia / etiology
  • Hypoglycemia / prevention & control
  • Hypoglycemic Agents / administration & dosage*
  • Insulin / administration & dosage*
  • Insulin Infusion Systems*
  • Male
  • Pilot Projects
  • Skiing / physiology

Substances

  • Blood Glucose
  • Glycated Hemoglobin A
  • Hypoglycemic Agents
  • Insulin
  • hemoglobin A1c protein, human