Real-Time US-18FDG-PET/CT Image Fusion for Guidance of Thermal Ablation of 18FDG-PET-Positive Liver Metastases: The Added Value of Contrast Enhancement

Cardiovasc Intervent Radiol. 2019 Jan;42(1):60-68. doi: 10.1007/s00270-018-2082-1. Epub 2018 Oct 4.

Abstract

Purpose: To assess the feasibility of US-18FDG-PET/CT fusion-guided microwave ablation of liver metastases either poorly visible or totally undetectable with US, CEUS and CT, but visualized by PET imaging.

Materials and methods: Twenty-three patients with 58 liver metastases underwent microwave ablation guided by image fusion system that combines US with 18FDG-PET/CT images. In 28/58 tumors, 18FDG-PET/CT with contrast medium (PET/CECT) was used. The registration technical feasibility, registration time, rates of correct targeting, technical success at 24 h, final result at 1 year and complications were analyzed and compared between the PET/CT and PET/CECT groups.

Results: Registration was successfully performed in all cases with a mean time of 7.8 + 1.7 min (mean + standard deviation), (4.6 + 1.5 min for PET/CECT group versus 10.9 + 1.8 min for PET/CT group, P < 0.01). In total, 46/58 (79.3%) tumors were correctly targeted, while 3/28 (10.7%) and 9/30 (30%) were incorrectly targeted in PET/CT and PET/CECT group, respectively (P < 0.05). Complete ablation was obtained at 24 h in 70.0% of cases (n = 40 tumors), 23/28 (82.1%) in the PET/CECT group and 17/30 (56.7%) in the PET/CT group (P < 0.037). Fourteen tumors underwent local retreatment (11 ablations, 2 with resection and 1 with stereotactic body radiation therapy), while 4 tumors could not be retreated because of distant disease progression and underwent systemic therapy. Finally, 54/58 (93.1%) tumors were completely treated at 1 year. One major complication occurred, a gastrointestinal hemorrhage which required surgical repair.

Conclusions: Percutaneous ablation of 18FDG-PET-positive liver metastases using fusion imaging of real-time US and pre-acquired 18FDG-PET/CT images is feasible, safe and effective. Contrast-enhanced PET/CT improves overall ablation accuracy and shortens procedural duration time.

Keywords: Fusion imaging; Liver; Metastasis; PET; Thermal ablation; Ultrasound.

MeSH terms

  • Aged
  • Contrast Media*
  • Feasibility Studies
  • Female
  • Fluorodeoxyglucose F18
  • Humans
  • Liver Neoplasms / diagnostic imaging
  • Liver Neoplasms / secondary*
  • Liver Neoplasms / surgery*
  • Male
  • Microwaves / therapeutic use
  • Middle Aged
  • Multimodal Imaging / methods*
  • Positron Emission Tomography Computed Tomography*
  • Prospective Studies
  • Surgery, Computer-Assisted / methods*
  • Ultrasonography, Interventional*

Substances

  • Contrast Media
  • Fluorodeoxyglucose F18