In-Clinic Evaluation of the MiniMed 670G System "Suspend Before Low" Feature in Children with Type 1 Diabetes

Diabetes Technol Ther. 2018 Nov;20(11):731-737. doi: 10.1089/dia.2018.0209. Epub 2018 Oct 6.

Abstract

Background: The Medtronic predictive low-glucose management (PLGM) algorithm automatically stops insulin delivery when sensor glucose (SG) is predicted to reach or fall below a preset low-glucose value within the next 30 min, and resumes delivery after hypoglycemia recovery. The present study evaluated the PLGM algorithm performance of the MiniMed™ 670G system SmartGuard™ "suspend before low" feature in children aged 7-13 years with type 1 diabetes (T1D).

Method: Participants (N = 105, mean ± standard deviation of 10.8 ± 1.8 years) underwent an overnight in-clinic evaluation of the "suspend before low" feature with a preset low limit of 65 mg/dL. After exercise, frequent sample testing (FST) was conducted every 5 min if values were <70 mg/dL; every 15 min if 70-80 mg/dL; and every 30 min if >80 mg/dL. First-day performance of the Guardian™ Sensor 3 glucose sensor and continuous glucose monitoring system was also evaluated.

Results: Activation of the "suspend before low" feature occurred in 79 of the 105 participants, 79.7% (63/79) did not result in SG falling below 65 mg/dL. Mean glucose at activation was 102 ± 19 mg/dL and the initial insulin suspension duration was 87.5 ± 32.7 min. Four hours after insulin resumption, mean reference glucose was 130 ± 42 mg/dL. Mean absolute relative difference between the FST reference glucose and SG values on the first day of sensor wear was 11.4%. For the 26 participants in whom the "suspend before low" feature did not activate, none involved a reference glucose value ≤65 mg/dL, suggesting that the PLGM algorithm performed as intended.

Conclusion: In children aged 7-13 years with T1D, the "suspend before low" feature of the MiniMed 670G system demonstrated a hypoglycemia prevention rate of nearly 80% after exercise and did not involve rebound hyperglycemia. There were no events of severe hypoglycemia during the evaluation.

Keywords: Automated insulin delivery.; Hybrid closed loop; Hypoglycemia; Insulin pump suspension; Pediatrics; Predictive low glucose management; Type 1 diabetes.

Publication types

  • Evaluation Study

MeSH terms

  • Adolescent
  • Algorithms*
  • Blood Glucose / analysis*
  • Blood Glucose Self-Monitoring / instrumentation
  • Blood Glucose Self-Monitoring / methods
  • Child
  • Diabetes Mellitus, Type 1 / blood
  • Diabetes Mellitus, Type 1 / complications
  • Diabetes Mellitus, Type 1 / drug therapy*
  • Exercise / physiology
  • Female
  • Humans
  • Hyperglycemia / blood
  • Hyperglycemia / etiology
  • Hyperglycemia / prevention & control
  • Hypoglycemia / blood
  • Hypoglycemia / etiology
  • Hypoglycemia / prevention & control*
  • Hypoglycemic Agents / administration & dosage*
  • Insulin / administration & dosage*
  • Insulin / analysis
  • Insulin Infusion Systems*
  • Male
  • Reference Values
  • Time Factors
  • Treatment Outcome