Feasibility of functional MRI at ultralow magnetic field via changes in cerebral blood volume

Neuroimage. 2019 Feb 1:186:185-191. doi: 10.1016/j.neuroimage.2018.10.071. Epub 2018 Oct 28.

Abstract

We investigate the feasibility of performing functional MRI (fMRI) at ultralow field (ULF) with a Superconducting QUantum Interference Device (SQUID), as used for detecting magnetoencephalography (MEG) signals from the human head. While there is negligible magnetic susceptibility variation to produce blood oxygenation level-dependent (BOLD) contrast at ULF, changes in cerebral blood volume (CBV) may be a sensitive mechanism for fMRI given the five-fold spread in spin-lattice relaxation time (T1) values across the constituents of the human brain. We undertook simulations of functional signal strength for a simplified brain model involving activation of a primary cortical region in a manner consistent with a blocked task experiment. Our simulations involve measured values of T1 at ULF and experimental parameters for the performance of an upgraded ULFMRI scanner. Under ideal experimental conditions we predict a functional signal-to-noise ratio of between 3.1 and 7.1 for an imaging time of 30 min, or between 1.5 and 3.5 for a blocked task experiment lasting 7.5 min. Our simulations suggest it may be feasible to perform fMRI using a ULFMRI system designed to perform MRI and MEG in situ.

Keywords: Cerebral blood volume; SQUID; Ultralow field; fMRI.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Brain / diagnostic imaging*
  • Cerebral Blood Volume*
  • Feasibility Studies
  • Functional Neuroimaging / instrumentation
  • Functional Neuroimaging / standards*
  • Humans
  • Image Processing, Computer-Assisted / standards*
  • Magnetic Fields
  • Magnetic Resonance Imaging*
  • Models, Theoretical*