Ultra-small-angle X-ray photon correlation spectroscopy using the Eiger detector

J Synchrotron Radiat. 2018 Nov 1;25(Pt 6):1753-1759. doi: 10.1107/S1600577518013899. Epub 2018 Oct 24.

Abstract

Successful implementation of the single-photon-counting Eiger 500k pixel array detector for sub-millisecond X-ray photon correlation spectroscopy (XPCS) measurements in the ultra-small-angle scattering region is reported. The performance is demonstrated by measuring the dynamics of dilute silica colloids in aqueous solvents when the detector is operated at different counter depths, 4, 8 and 12 bit. In the fastest mode involving 4 bit parallel readout, a stable frame rate of 22 kHz is obtained that enabled measurement of intensity-intensity autocorrelation functions with good statistics down to the 50 µs range for a sample with sufficient scattering power. The high frame rate and spatial resolution together with large number of pixels of the detector facilitate the investigation of sub-millisecond dynamics over a broad length scale by multispeckle XPCS. This is illustrated by an example involving phoretic motion of colloids during the phase separation of the solvent.

Keywords: USAXS; XPCS; coherent X-ray scattering; colloid dynamics.