Analytical and Clinical Performance Evaluation of the Elecsys HIV combi PT Assay on the cobas e 602 Analyzer for the Diagnosis of Human Immunodeficiency Virus

Am J Clin Pathol. 2019 Mar 1;151(4):377-385. doi: 10.1093/ajcp/aqy153.

Abstract

Objectives: We evaluated the performance of the Elecsys HIV combi PT assay on the cobas e 602 analyzer for diagnosing human immunodeficiency virus (HIV; part of the US Food and Drug Administration [FDA] submission).

Methods: The HIV combi PT and reference (ARCHITECT HIV Ag/Ab Combo) assays were assessed at four independent clinical laboratories/one reference laboratory (United States; July 2014 to November 2015). Clinical performance was evaluated using four reagent lots. Analytical performance was evaluated per Clinical and Laboratory Standards Institute EP05-A3 guidelines. Serum/plasma samples from 18 clinical sites/vendors (United States and outside the United States) were tested.

Results: Sensitivity (95% confidence interval [CI]) in HIV-1 antibody-positive individuals (United States and outside the United States; n = 1,460) was 100.00% (99.75%-100.00%). Specificity was 99.94% (95% CI, 99.85%-99.98%) in low-risk individuals (United States; n = 6,843), 98.19% (95% CI, 96.93%-99.04%) in high-risk individuals (United States and outside the United States; n = 758), and 97.43% (95% CI, 95.32%-98.76%) in pregnant women (United States and outside the United States; n = 440). Analytical performance was acceptable.

Conclusions: We demonstrate the robustness of the FDA-approved Elecsys HIV combi PT assay on the cobas e 602 analyzer for HIV testing in the United States.

Keywords: HIV immunoassay; HIV p24 antigen; HIV-1; HIV-2; Sensitivity; Specificity.

Publication types

  • Evaluation Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Diagnosis-Related Groups
  • Female
  • HIV Antibodies / blood*
  • HIV Infections / blood
  • HIV Infections / diagnosis*
  • HIV Infections / immunology
  • HIV Infections / virology
  • HIV-1 / immunology*
  • HIV-1 / isolation & purification
  • Humans
  • Length of Stay
  • Mass Screening / instrumentation
  • Mass Screening / methods
  • Pregnancy
  • Prothrombin Time
  • Sensitivity and Specificity
  • United States

Substances

  • HIV Antibodies