Aim: Dysfunctional arteriovenous (AV) access remains a significant cause of morbidity and hospital admission for patients with end stage renal failure on haemodialysis. This study was performed to evaluate the impact of paclitaxel-coated Balloon (PCB) on the patency of AV access with recurrent stenoses.
Methods: We retrospectively studied haemodialysis patients who presented to our centre with recurrent AV access dysfunction and compared intervention-free patency using plain balloon versus PCB.
Results: A total of 147 patients were followed up longitudinally. Intervention-free patency was better following PCB compared to previous intervention using plain balloons (6.4 ± 5.8 versus 4.0 ± 3.7, P < 0.01). The 3- and 6-month patency rates after PCB were significantly better compared to standard plain angioplasty balloon: 69.4% versus 52.4%, P < 0.01 and 42.9% versus 15.6%, P < 0.01 respectively. Kaplan-Meier survival analysis of circuit patency demonstrated the superiority of PCB over plain balloon angioplasty in both arteriovenous fistula and arteriovenous graft (P < 0.01 and P = 0.01 respectively) although the patency of arteriovenous fistula remained significantly better than arteriovenous graft following interventions with PCB (P < 0.01). Age of AV access and the number of previous interventions were found to be significant predictors of patency following PCB intervention.
Conclusion: Arteriovenous access intervention with PCB was shown to be superior compared to plain balloon in the treatment of both non-thrombosed and thrombosed AV accesses in our multi-ethnic population.