Validation of a novel smartphone application-enabled, patient-operated skin barrier device

Skin Res Technol. 2019 Sep;25(5):612-617. doi: 10.1111/srt.12692. Epub 2019 Apr 3.

Abstract

Background: Transepidermal water loss (TEWL) and surface capacitance measure skin barrier permeability and stratum corneum (SC) hydration, respectively, and are frequently utilized in atopic dermatitis clinical trials. Many barrier devices are costly and often used only in the academic setting. GPSkin is a low-cost, patient-operated device that measures both TEWL and SC hydration. This study aimed to test the reliability of GPSkin and assess its correlation with current industry standards.

Materials and methods: GPSkin was compared to the Biox AquaFlux (TEWL) and Courage-Khazaka Corneometer (SC hydration). Participants with healthy skin (n = 50) collected measurements with GPSkin in Trial 1 without any device education and in Trial 2 with additional instruction. In Trial 2, the investigator also performed measurements with GPSkin. Spearman's coefficients (rs ) were performed to assess device correlation. Intraclass correlation coefficients (ICC) were calculated to determine reliability.

Results: Overall, GPSkin was moderately correlated with current industry device measurements for TEWL (Trial 1 rs :0.48; Trial 2 rs :0.40 participant, 0.34 investigator) and SC hydration (Trial 1 rs :0.63; Trial 2 rs :0.45). GPSkin demonstrated "good" test-retest reliability for both TEWL (ICC: 0.89) and SC hydration (ICC: 0.85) measurements when participants were provided with some device education. There was no difference in reliability between participants provided with device education and investigators.

Conclusion: Based on these findings, we concluded that GPSkin provides reasonably precise and reliable measurements of SC hydration and TEWL as compared to current devices.

Keywords: atopic dermatitis; barrier; capacitance; device; hydration; transepidermal water loss.

Publication types

  • Validation Study

MeSH terms

  • Adolescent
  • Adult
  • Aged
  • Biosensing Techniques / instrumentation
  • Biosensing Techniques / methods
  • Cross-Sectional Studies
  • Epidermis / physiology*
  • Equipment Design
  • Humans
  • Middle Aged
  • Mobile Applications*
  • Organism Hydration Status / physiology
  • Permeability
  • Prospective Studies
  • Reproducibility of Results
  • Skin Physiological Phenomena
  • Smartphone*
  • Water Loss, Insensible / physiology*
  • Young Adult

Grants and funding