Background: The Onvision needle tip tracking (NTT) is a new technology consisting of a needle with an ultrasound sensor close to the needle tip and a console for computerised signal processing. The aim of the study was to evaluate NTT technology during ultrasound-guided simulated peripheral nerve block procedures in a porcine phantom model.
Methods: Forty anaesthesiologists performed in-plane and out-of-plane simulated nerve blocks with and without NTT guidance. The primary outcome measure was procedure time. Secondary outcomes were hand movements and the path length travelled by the hands measured by motion analysis, precision of the needle tip related to the target structure, success rates and violations of the target structure, and the participants confidence whether their procedure would be successful or not.
Results: Procedure time was reduced from 66.7 (SD = 47.5) seconds to 43.8 (SD = 29.2) seconds when NTT was used for out-of-plane procedures (P = 0.002). The number of hand movements of the probe hand was 13.9 (SD = 30.2) with NTT and 22.8 (SD = 30.0) without NTT (P = 0.019). No significant differences were registered during the performance of in-plane procedures. The participants confidence in a presumed block success was increased with both in-plane procedures (8.50 (SD = 1.18) with NTT vs 7.65 (SD = 1.96), P = 0.004) and out-of-plane procedures (8.50 (SD = 1.09) vs 7.10 (SD = 1.89), P = 0.0001).
Conclusions: The new NTT technology significantly reduced the procedure time and the number of hand movements for ultrasound-guided out-of-plane PNB procedures. No significant differences were found for the in-plane procedures.
© 2019 The Authors. Acta Anaesthesiologica Scandinavica Foundation published by John Wiley & Sons Ltd on behalf of Acta Anaesthesiologica Scandinavica Foundation.