Use of sensor-integrated pump therapy to reduce hypoglycaemia in people with Type 1 diabetes: a real-world study in the UK

Diabet Med. 2019 Sep;36(9):1100-1108. doi: 10.1111/dme.14043. Epub 2019 Jun 30.

Abstract

Aims: To assess the efficacy of insulin pumps with automated insulin suspension systems in a real-world setting.

Methods: We analysed anonymized data uploaded to CareLink by people (n=920) with Type 1 diabetes using the MiniMed Paradigm Veo system and the MiniMed 640G system (Medtronic International Trading Sàrl, Tolochanez, Switzerland) with SmartGuard technology, with or without automated insulin suspension enabled, between February 2016 and June 2018. Users with ≥15 days of sensor data and ≥70% sensor-wear time were classified as sensor-augmented pump alone, sensor-integrated pump with low glucose suspend enabled or sensor-integrated pump with predictive low glucose management enabled.

Results: The median (25th -75th percentile) system use was 161 (58-348) days. The median time spent with sensor glucose values ≤3 mmol/l was 0.8 (0.3-1.7)% in the sensor-augmented pump group, 0.3 (0.1-0.7)% in the sensor-integrated pump with low glucose suspend group, and 0.3 (0.1-0.5)% in the sensor-integrated pump with predictive low glucose management group. In individuals switching from sensor-augmented pump to sensor-integrated pump with low glucose suspend (n=31), there were significant reductions in the monthly rate of hypoglycaemic events <3 mmol/l (rate ratio 0.63, 95% CI 0.45-0.89; P=0.009) and in the percentage of time with glucose values ≤3 mmol/l [sensor-augmented pump: 0.63% (95% CI 0.34-1.29), sensor-integrated pump with low glucose suspend: 0.33% (95% CI 0.16-0.64); P=0.001]. The monthly rate of hypoglycaemic events decreased further in individuals (n=139) switching from sensor-integrated pump with low glucose suspend to sensor-integrated pump with predictive low glucose management [rate ratio 0.82 (95% CI 0.69-0.98); P<0.0274]. Similar results were seen for events <3.9 mmol/l. There was no difference in median time spent in target glucose range.

Conclusion: Real-world UK data show that increasing automation of insulin suspension reduces hypoglycaemia exposure in people with Type 1 diabetes.

Publication types

  • Observational Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adolescent
  • Adult
  • Biosensing Techniques / instrumentation
  • Blood Glucose / analysis*
  • Blood Glucose Self-Monitoring / instrumentation
  • Child
  • Databases, Factual
  • Diabetes Mellitus, Type 1 / blood*
  • Diabetes Mellitus, Type 1 / complications
  • Diabetes Mellitus, Type 1 / drug therapy*
  • Diabetes Mellitus, Type 1 / epidemiology
  • Equipment Design
  • Female
  • Glycated Hemoglobin / analysis
  • Humans
  • Hypoglycemia / chemically induced
  • Hypoglycemia / epidemiology
  • Hypoglycemia / prevention & control*
  • Insulin / administration & dosage*
  • Insulin / adverse effects
  • Insulin Infusion Systems*
  • Male
  • Medical Record Linkage / methods
  • Time Factors
  • Treatment Outcome
  • United Kingdom / epidemiology
  • Young Adult

Substances

  • Blood Glucose
  • Glycated Hemoglobin A
  • Insulin