Objectives: This study sought to compare patients with and without long-standing persistent atrial fibrillation (LSPAF) undergoing Watchman left atrial appendage (LAA) occlusion.
Background: An increased burden of atrial fibrillation is associated with progressive left atrial remodeling and enlargement.
Methods: Transesophageal echocardiography (TEE) measures of LAA ostial diameter and depth, device compression, and residual leak were evaluated in 101 consecutive Watchman cases. The patients were categorized into LSPAF (n = 48) or non-LSPAF (n = 53) groups and compared.
Results: The average LAA ostial diameter for LSPAF versus non-LSPAF by TEE omniplane at 0° was 21.1 ± 4.1 mm versus 18.2 ± 3.6 mm (p = 0.0002); at 45° was 18.7 ± 3.4 mm versus 16.3 ± 3.1 mm (p = 0.0004); at 90° was 19.6 ± 3.8 mm versus 16.2 ± 3.4 mm (p = 0.00001); and at 135° was 21.0 ± 4.1 mm versus 18.0 ± 4.1 mm (p = 0.0005). The average LAA depth for LSPAF versus non-LSPAF by TEE at 0° was 28.1 ± 6.4 mm versus 25.2 ± 4.9 mm (p = 0.02); at 45° was 27.9 ± 5.8 mm versus 25.1 ± 4.3 mm (p = 0.007); at 90° was 27.2 ± 5.2 mm versus 22.8 ± 3.7 mm (p = 0.0001); and at 135° was 25.6 ± 5.4 mm versus 21.5 ± 3.8 mm (p = 0.0001). In successfully treated patients, 77% of the LSPAF group received larger device (27, 30, or 33 mm) implants versus only 46% in the non-LSPAF group (p = 0.003). While both groups had similar rates of moderate (3 to 5 mm) leaks at implant (2% vs. 0%; p = 0.14), 27% of the LSPAF vs. 4% of the non-LSPAF group had moderate leaks (p = 0.04) on 6-week follow-up TEE.
Conclusions: Patients with LSPAF have significantly larger LAA sizes, require larger devices, and have more residual leak on follow-up TEE. LSPAF may represent a higher risk group that warrants more stringent long-term follow-up.
Keywords: Watchman device; left atrial appendage occlusion; long-standing persistent atrial fibrillation; peridevice leak.
Copyright © 2019 American College of Cardiology Foundation. Published by Elsevier Inc. All rights reserved.