Risk factors and prognostic impact of left ventricular assist device-associated infections

Am Heart J. 2019 Aug:214:69-76. doi: 10.1016/j.ahj.2019.04.021. Epub 2019 May 6.

Abstract

Background: Left ventricular assist device (LVAD)-associated infections may be life-threatening and impact patients' outcome. We aimed to identify the characteristics, risk factors, and prognosis of LVAD-associated infections.

Methods: Patients included in the ASSIST-ICD study (19 centers) were enrolled. The main outcome was the occurrence of LVAD-associated infection (driveline infection, pocket infection, or pump/cannula infection) during follow-up.

Results: Of the 652 patients enrolled, 201 (30.1%) presented a total of 248 LVAD infections diagnosed 6.5 months after implantation, including 171 (26.2%), 51 (7.8%), and 26 (4.0%) percutaneous driveline infection, pocket infection, or pump/cannula infection, respectively. Patients with infections were aged 58.7 years, and most received HeartMate II (82.1%) or HeartWare (13.4%). Most patients (62%) had implantable cardioverter-defibrillators (ICDs) before LVAD, and 104 (16.0%) had ICD implantation, extraction, or replacement after the LVAD surgery. Main pathogens found among the 248 infections were Staphylococcus aureus (n = 113' 45.4%), Enterobacteriaceae (n = 61; 24.6%), Pseudomonas aeruginosa (n = 34; 13.7%), coagulase-negative staphylococci (n = 13; 5.2%), and Candida species (n = 13; 5.2%). In multivariable analysis, HeartMate II (subhazard ratio, 1.56; 95% CI, 1.03 to 2.36; P = .031) and ICD-related procedures post-LVAD (subhazard ratio, 1.43; 95% CI, 1.03-1.98; P = .031) were significantly associated with LVAD infections. Infections had no detrimental impact on survival.

Conclusions: Left ventricular assist device-associated infections affect one-third of LVAD recipients, mostly related to skin pathogens and gram-negative bacilli, with increased risk with HeartMate II as compared with HeartWare, and in patients who required ICD-related procedures post-LVAD. This is a plea to better select patients needing ICD implantation/replacement after LVAD implantation.

Trial registration: ClinicalTrials.gov NCT02873169.

Publication types

  • Multicenter Study
  • Observational Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Catheter-Related Infections / epidemiology
  • Catheter-Related Infections / etiology*
  • Catheter-Related Infections / microbiology
  • Catheter-Related Infections / mortality
  • Defibrillators, Implantable / statistics & numerical data
  • Device Removal / statistics & numerical data
  • Female
  • France / epidemiology
  • Heart Ventricles
  • Heart-Assist Devices / adverse effects*
  • Heart-Assist Devices / statistics & numerical data
  • Humans
  • Male
  • Middle Aged
  • Multivariate Analysis
  • Prognosis
  • Prosthesis-Related Infections / epidemiology
  • Prosthesis-Related Infections / etiology*
  • Prosthesis-Related Infections / microbiology
  • Prosthesis-Related Infections / mortality
  • Retrospective Studies
  • Risk Factors

Associated data

  • ClinicalTrials.gov/NCT02873169