Continuous intracoronary versus standard intravenous infusion of adenosine for fractional flow reserve assessment: the HYPEREMIC trial

EuroIntervention. 2020 Sep 18;16(7):560-567. doi: 10.4244/EIJ-D-18-01067.

Abstract

Aims: The aim of this study was to evaluate the accuracy of a continuous intracoronary (IC) adenosine infusion, administered through the novel HYPEREM™IC over-the-wire microcatheter, to measure fractional flow reserve (FFR).

Methods and results: The HYPEREMIC trial was a randomised, non-inferiority, crossover study in which patients with intermediate coronary lesions were enrolled for sequential pressure wire studies. FFR was measured using intravenous (IV) (140-180 mcg/kg/min) versus continuous non-weight-adjusted IC (360 mcg/min) adenosine. Patients were randomised and blinded to the order in which they received the adenosine, separated by a washout period. The primary endpoint was the mean hyperaemic FFR. Forty-one patients were enrolled at three UK sites between June and November 2016. The mean (standard deviation) FFR was 0.82 (±0.09) after IC versus 0.84 (±0.09) after IV adenosine. The difference of -0.02 (95% confidence interval [CI]: -0.03 to -0.01) confirmed the non-inferiority (margin <0.05) of IC to IV adenosine. Intracoronary adenosine was associated with a shorter mean time to maximal hyperaemia (difference -44 [95% CI: -59 to -29] seconds; p<0.0001). Chest discomfort was reported in 32/41 (78.0%) patients during IV adenosine versus 12/41 (29.3%) patients during IC adenosine.

Conclusions: Continuous IC adenosine was a reliable, faster and better tolerated method of achieving maximal hyperaemia compared to IV adenosine.

MeSH terms

  • Adenosine
  • Coronary Angiography
  • Coronary Stenosis / diagnosis*
  • Coronary Vessels / diagnostic imaging
  • Cross-Over Studies
  • Fractional Flow Reserve, Myocardial / drug effects*
  • Humans
  • Hyperemia*
  • Infusions, Intravenous
  • Vasodilator Agents / pharmacology

Substances

  • Vasodilator Agents
  • Adenosine