Low-Profile Intra-Aneurysmal Flow Disruptor WEB 17 versus WEB Predecessor Systems for Treatment of Small Intracranial Aneurysms: Comparative Analysis of Procedural Safety and Feasibility

AJNR Am J Neuroradiol. 2019 Oct;40(10):1766-1772. doi: 10.3174/ajnr.A6183. Epub 2019 Sep 5.

Abstract

Background and purpose: The Woven EndoBridge 17 has recently been introduced to the market for facilitated endovascular treatment of small bifurcation aneurysms (≤7 mm) with low-profile microcatheters. We compared the Woven EndoBridge 17 with its predecessor versions in terms of procedural safety and feasibility.

Materials and methods: This was a multicenter review of aneurysms ranging from 3 to 7 mm treated with the Woven EndoBridge between 2011 and 2019. Aneurysm characteristics, procedural parameters, and complications were retrospectively compared between treatment with the Woven EndoBridge 17 and a control group that was treated with its predecessor versions, using inverse probability of treatment weighting.

Results: Thirty-eight aneurysms treated with a Woven EndoBridge 17 (mean size, 4.9 ± 1.5 mm) and 70 treated with a predecessor version of the Woven EndoBridge 17 (mean size, 5.6 ± 1.4 mm) were included. The predecessor version of the Woven EndoBridge 17 had a higher failure rate (10.3%) than the Woven EndoBridge 17 (0%, P = .05). Additional stent placement was performed more often with the predecessor version of the Woven EndoBridge 17 (10.0%) than with the Woven EndoBridge 17 (2.6%, adjusted P = .005). The predecessor version of the Woven EndoBridge 17 was associated with a higher thromboembolic event rate (14.3%) than the Woven EndoBridge 17 (5.3%, adjusted P = .002). Neurologic complications (Woven EndoBridge 17: 2.6%; predecessor version of the Woven EndoBridge 17: 2.9%, adjusted P = 1.0) and immediate complete aneurysm occlusion rates (Woven EndoBridge 17: 57.9%; predecessor version of the Woven EndoBridge 17: 54.3%, adjusted P = .21) did not differ significantly between groups.

Conclusions: In the current study, the Woven EndoBridge 17 was associated with a potentially lower thromboembolic event rate than the predecessor version of the Woven EndoBridge 17, without compromising the immediate aneurysm occlusion rate. Long-term clinical and angiographic outcome analysis will be necessary to draw a definite conclusion.

Publication types

  • Comparative Study
  • Multicenter Study

MeSH terms

  • Adult
  • Aged
  • Aged, 80 and over
  • Blood Vessel Prosthesis* / adverse effects
  • Embolization, Therapeutic / adverse effects
  • Embolization, Therapeutic / instrumentation*
  • Embolization, Therapeutic / methods
  • Endovascular Procedures / adverse effects
  • Endovascular Procedures / instrumentation*
  • Endovascular Procedures / methods
  • Female
  • Humans
  • Intracranial Aneurysm / therapy*
  • Male
  • Middle Aged
  • Postoperative Complications / epidemiology
  • Postoperative Complications / etiology*
  • Retrospective Studies
  • Treatment Outcome