Implantation of one versus two MitraClips in the German TRAMI registry: Is more always better?

Catheter Cardiovasc Interv. 2020 Sep 1;96(3):E360-E368. doi: 10.1002/ccd.28613. Epub 2019 Dec 2.

Abstract

Objectives: To compare baseline characteristics and outcomes in patients treated with either 1 or 2 MitraClips in the German TRAMI (Transcatheter Mitral Valve Interventions) registry.

Background: The MitraClip community seems to silently assume that results should intrinsically be better after implantation of more than one clip, although data is still sparse.

Methods: In 2010-2013, 803 patients were enrolled prospectively into TRAMI (461 one-clip and 312 two-clip procedures). Follow-up was performed centrally at 30 days and 1 year.

Results: Baseline characteristics of TRAMI-patients with two clips differed significantly from single-clip patients regarding constitutional (more men, taller body height) and heart failure-related factors (larger left ventricular dimensions, reduced left ventricular ejection fraction, more severe heart failure). Also, a significant increase in two-clip procedures over time was present. After propensity score matching for differing baseline characteristics, residual moderate mitral regurgitation (MR) occurred more frequently after implantation of two clips, whereas residual severe MR could more frequently be observed after one-clip procedures. However, no or mild residual MR at discharge was present in 71.6% after single-clip and in 70.1% after two-clips implantation (p = .81). After 1 year, no significant differences regarding mortality or New York Heart Association status could be detected in the propensity matched cohorts. However, TRAMI-patients treated with two clips had a significantly higher incidence of cerebral-vascular events (p = .02).

Conclusions: TRAMI data cannot support the theory that implantation of more than one clip is associated with better clinical outcomes. The finding of more cerebral-vascular events after two-clip procedures might be hypothesis-generating.

Keywords: MitraClip; TRAMI registry; cerebral-embolic risk; mitral regurgitation; percutaneous edge-to-edge mitral valve repair; single and dual clip implantation; stroke.

Publication types

  • Comparative Study

MeSH terms

  • Aged
  • Aged, 80 and over
  • Female
  • Germany
  • Heart Valve Prosthesis Implantation / adverse effects
  • Heart Valve Prosthesis Implantation / instrumentation*
  • Heart Valve Prosthesis Implantation / mortality
  • Heart Valve Prosthesis*
  • Hemodynamics
  • Humans
  • Male
  • Mitral Valve / diagnostic imaging
  • Mitral Valve / physiopathology
  • Mitral Valve / surgery*
  • Mitral Valve Annuloplasty / adverse effects
  • Mitral Valve Annuloplasty / instrumentation*
  • Mitral Valve Annuloplasty / mortality
  • Mitral Valve Insufficiency / diagnostic imaging
  • Mitral Valve Insufficiency / mortality
  • Mitral Valve Insufficiency / physiopathology
  • Mitral Valve Insufficiency / surgery*
  • Postoperative Complications / mortality
  • Prospective Studies
  • Prosthesis Design
  • Recovery of Function
  • Registries
  • Risk Assessment
  • Risk Factors
  • Time Factors
  • Treatment Outcome