Transesophageal echocardiography complications associated with interventional cardiology procedures

Am Heart J. 2020 Mar:221:19-28. doi: 10.1016/j.ahj.2019.11.018. Epub 2019 Dec 9.

Abstract

Background: Although there have been several reports documenting complications related with transesophageal echocardiography (TEE) manipulation following cardiac surgery, there is a paucity of data regarding the safety of TEE used to guide catheter-based interventions. The aim of this study was to determine the prevalence, types and risk factors of complications associated with procedures requiring active TEE guidance.

Methods: This study included 1249 consecutive patients undergoing either transcatheter aortic valve implantation (TAVI), Mitraclip, left atrial appendage occlusion (LAAO) or paravalvular leak closure (PVLC). Patients were divided into 2 cohorts based on the degree of probe manipulation required to guide the procedure and the risk of developing a TEE-related complication: low-risk (TAVI, n = 1037) and high-risk (Mitraclip, LAAO and PVLC, n = 212). Patients were further analyzed according to the occurrence of major and minor TEE-related complications.

Results: The overall incidence of TEE-related complications was 0.9% in the TAVI group and 6.1% in the rest of the cohort (P < .001). Patients in the high-risk cohort had also a higher incidence of major-complications (2.8% vs 0.6%, P = .008), and factors associated with an increased risk were being underweight, having a prior history of gastrointestinal bleeding and the use of chronic steroids/immunosuppressive medications. Procedural time under TEE-manipulation was longer in patients exhibiting complications and was an independent predictor of major complications (OR = 1.13, 95% CI 1.01-1.25, for each 10 minutes increments in imaging time). Patients with major complications undergoing Mitraclip had the longest median time under TEE-manipulation (297 minutes) and a risk of developing a major-complication that was 10.64 times higher than the rest of the cohort (95% CI 3.30-34.29, P < .001).

Conclusion: The prevalence of TEE-related complications associated with interventional procedures is higher than previously reported. Undergoing a prolonged procedure, particularly in the setting of Mitraclip, was the main factor linked to TEE-related complications.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Aged
  • Aged, 80 and over
  • Atrial Appendage / surgery
  • Blood Transfusion
  • Cardiac Catheterization / methods*
  • Deglutition Disorders / epidemiology*
  • Echocardiography, Transesophageal / adverse effects*
  • Esophageal Perforation / epidemiology
  • Esophageal Perforation / etiology
  • Esophagus / injuries*
  • Female
  • Gastrointestinal Hemorrhage / epidemiology*
  • Gastrointestinal Hemorrhage / etiology
  • Gastrointestinal Hemorrhage / therapy
  • Humans
  • Lacerations / epidemiology
  • Lacerations / etiology
  • Male
  • Mitral Valve / surgery
  • Postoperative Complications / epidemiology*
  • Postoperative Complications / etiology
  • Retrospective Studies
  • Surgery, Computer-Assisted
  • Time Factors
  • Transcatheter Aortic Valve Replacement / methods*