First-in-man study of a thinner-strut sirolimus-eluting bioresorbable scaffold (FUTURE-I): Three-year clinical and imaging outcomes

Catheter Cardiovasc Interv. 2020 Feb:95 Suppl 1:648-657. doi: 10.1002/ccd.28722. Epub 2020 Jan 7.

Abstract

Objectives: The FUTURE-I study aimed to assess preliminary safety and effectiveness with the long-term clinical and imaging follow-up for the Firesorb (MicroPort, Shanghai, China), a thinner-strut sirolimus-eluting bioresorbable scaffold (BRS).

Background: First-generation BRS has been associated with unexpected device-related adverse outcomes at long-term follow-up.

Methods: In this prospective, open-label, first-in-man study, patients with single de novo lesions in native coronary arteries were randomized 2:1 into two cohorts after successful Firesorb implantation: cohort 1 (n = 30) underwent multimodality imaging assessment at 6 and 24 months; and cohort 2 (n = 15) at 12 and 36 months. All patients underwent clinical follow-up at 1, 6, and 12 months and annually up to 5 years.

Results: Between January and March 2016, 45 patients were enrolled. At 3-year follow-up, one patient had experienced target lesion failure and none scaffold thrombosis. In-scaffold minimal lumen diameter decreased significantly from 6-month to 2-year (2.53 ± 0.24 mm vs. 2.27 ± 0.37 mm, p = .0003), and only numerically from 1-year to 3-year follow-up (2.48 ± 0.28 mm vs. 2.22 ± 0.13 mm, p = .08). By optical coherence tomography, neointimal strut coverage at 3-year follow-up was 99.8%, and very low rate of late scaffold discontinuity was observed, only in one patient on two cross sections with three malapposed struts.

Conclusions: At 3-year follow-up of the FUTURE-I study, implantation of the thinner-strut Firesorb BRS appeared preliminary feasible and effective in the treatment of patients with noncomplex coronary lesions.

Keywords: bioresorbable scaffold; intravascular ultrasound; optical coherence tomography.

Publication types

  • Randomized Controlled Trial

MeSH terms

  • Absorbable Implants*
  • Angioplasty, Balloon, Coronary / adverse effects
  • Angioplasty, Balloon, Coronary / instrumentation*
  • Cardiovascular Agents / administration & dosage*
  • Cardiovascular Agents / adverse effects
  • China
  • Coated Materials, Biocompatible*
  • Coronary Angiography
  • Coronary Artery Disease / diagnostic imaging
  • Coronary Artery Disease / therapy*
  • Female
  • Humans
  • Male
  • Middle Aged
  • Prospective Studies
  • Prosthesis Design
  • Sirolimus / administration & dosage*
  • Sirolimus / adverse effects
  • Time Factors
  • Tomography, Optical Coherence
  • Treatment Outcome
  • Ultrasonography, Interventional

Substances

  • Cardiovascular Agents
  • Coated Materials, Biocompatible
  • Sirolimus