Abstract

This work proposes a new method for pseudo-3D verification of intensity-modulated radiation therapy (IMRT) dose distributions. Unlike commercial solutions, it uses measured doses only for gamma evaluation. Its resolution is far better than with electronic detectors within the measured plane and comparable in other directions. It is readily available at clinics because it uses existing resources-a slab phantom and EBT3 films. The method was tested on six IMRT clinical cases. An in-house code for 2D and pseudo-3D gamma analysis was written in MATLAB and compared to OmniPro I'mRT.