Two-year follow-up of a randomized multicenter study comparing a drug-coated balloon with a drug-eluting stent in native small coronary vessels: The RESTORE Small Vessel Disease China trial

Catheter Cardiovasc Interv. 2020 Feb:95 Suppl 1:587-597. doi: 10.1002/ccd.28705. Epub 2020 Jan 13.

Abstract

Objectives: To report the clinical outcomes of the RESTORE drug-coated balloon (DCB; Cardionovum, Bonn, Germany) for treatment of de novo small vessel disease (SVD) beyond 1 year.

Background: Previous reports have demonstrated the noninferiority of the RESTORE DCB to the RESOLUTE Integrity drug-eluting stent (DES; Medtronic, Minneapolis, Minnesota) in terms of 9-month in-segment percent diameter stenosis.

Methods: In the prospective, multicenter, noninferiority RESTORE SVD China trial, 230 patients with visually-estimated reference vessel diameter (RVD) ≥2.25 and ≤2.75 mm were randomized to DCB or DES in a 1:1 ratio stratified by diabetes and number of lesions treated. Furthermore, 32 patients with RVD ≥2.00 and <2.25 mm were enrolled in a nested very small vessel (VSV) registry. Clinical follow-up were performed at 2 years to evaluate target lesion failure (TLF) in both groups and the VSV cohort.

Results: Overall, 256 (97.7%) patients (115 and 109 in the DCB and DES groups, respectively, and 32 in the VSV cohort) completed 2 years of follow-up. There was no significant difference in TLF between the DCB and DES groups (5.2 vs. 3.7%, p = .75). Target lesion revascularization was acceptable at 1 month, 1 year, and 2 years, and did not differ significantly with DCB from that in the DES group (0.9 vs. 0%, p = 1.0, 4.4 vs. 2.6%, p = .72, 5.2 vs. 2.8%, p = .50, respectively).

Conclusions: Compared to the second-generation DES, the RESTORE DCB did not increase the risk of clinical outcomes. Late catch-up phenomen requiring revascularization was not significant in this study.

Keywords: angioplasty, balloon, coronary; coronary artery disease; follow-up studies; percutaneous coronary intervention; stents.

Publication types

  • Comparative Study
  • Equivalence Trial
  • Multicenter Study

MeSH terms

  • Aged
  • Angioplasty, Balloon, Coronary / adverse effects
  • Angioplasty, Balloon, Coronary / instrumentation*
  • Cardiac Catheters*
  • China
  • Coated Materials, Biocompatible*
  • Coronary Artery Disease / diagnostic imaging
  • Coronary Artery Disease / therapy*
  • Drug-Eluting Stents*
  • Female
  • Humans
  • Male
  • Middle Aged
  • Registries
  • Risk Assessment
  • Risk Factors
  • Time Factors
  • Treatment Outcome

Substances

  • Coated Materials, Biocompatible