Late onset of new conduction disturbances requiring permanent pacemaker implantation following TAVI

Heart. 2020 Aug;106(16):1244-1251. doi: 10.1136/heartjnl-2019-315967. Epub 2020 Jan 31.

Abstract

Background: The timing of onset and associated predictors of late new conduction disturbances (CDs) leading to permanent pacemaker implantation (PPI) following transcatheter aortic valve implantation (TAVI) are still unknown, however, essential for an early and safe discharge. This study aimed to investigate the timing of onset and associated predictors of late onset CDs in patients requiring PPI (LCP) following TAVI.

Methods and results: We performed retrospective analysis of prospectively collected data from five large volume centres in Europe. Post-TAVI electrocardiograms and telemetry data were evaluated in patients with a PPI post-TAVI to identify the onset of new advanced CDs. Early onset CDs were defined as within 48 hours after procedure, and late onset CDs as after 48 hours. A total of 2804 patients were included for analysis. The PPI rate was 12%, of which 18% was due to late onset CDs (>48 hours). Independent predictors for LCP were pre-existing non-specific intraventricular conduction delay, pre-existing right bundle branch block, self-expandable valves and predilation. At least one of these risk factors was present in 98% of patients with LCP. Patients with a balloon-expandable valve without predilation did not develop CDs requiring PPI after 48 hours.

Conclusions: Safe early discharge might be feasible in patients without CDs in the first 48 hours after TAVI if no risk factors for LCP are present.

Keywords: aortic stenosis; transcatheter valve interventions; valvular heart disease.

Publication types

  • Multicenter Study

MeSH terms

  • Aged
  • Aged, 80 and over
  • Aortic Valve / diagnostic imaging
  • Aortic Valve / physiopathology
  • Aortic Valve / surgery*
  • Aortic Valve Stenosis / diagnostic imaging
  • Aortic Valve Stenosis / mortality
  • Aortic Valve Stenosis / physiopathology
  • Aortic Valve Stenosis / surgery*
  • Arrhythmias, Cardiac / diagnosis
  • Arrhythmias, Cardiac / mortality
  • Arrhythmias, Cardiac / physiopathology
  • Arrhythmias, Cardiac / therapy*
  • Cardiac Pacing, Artificial* / adverse effects
  • Cardiac Pacing, Artificial* / mortality
  • Databases, Factual
  • Europe
  • Female
  • Humans
  • Length of Stay
  • Male
  • Pacemaker, Artificial*
  • Patient Discharge
  • Retrospective Studies
  • Risk Assessment
  • Risk Factors
  • Time Factors
  • Transcatheter Aortic Valve Replacement / adverse effects*
  • Transcatheter Aortic Valve Replacement / mortality
  • Treatment Outcome