Background: Treponema-specific assays are widely adopted in the first step of the reverse algorithm of serologic syphilis screening. The new BioPlex 2200 Syphilis Total and rapid plasma reagin (RPR) test is designed to perform the first 2 steps of the algorithm simultaneously. However, limited data regarding the BioPlex Syphilis Total and RPR in clinical practice exist.
Methods: A total of 293 random samples at a tertiary medical center were tested by BioPlex Syphilis Total and RPR, BioPlex Syphilis IgG, Architect Syphilis TP, and BD Macro-Vue RPR card. Treponema pallidum particle agglutination (TP-PA) assay and clinical chart review were used to resolve discrepancies. Comparisons were performed among treponemal-specific assays and between 2 RPR tests.
Results: Good overall agreements (>91%) were achieved between BioPlex Syphilis Total, BioPlex Syphilis IgG, and Architect Syphilis TP. Overall agreement between BioPlex RPR and BD RPR was 86.8% with positive percent agreement of 66.7% and negative percent agreement of 96.3%. There were 37 discordant samples including 30 with BD RPR+/BioPlex RPR- and 7 with BD RPR-/BioPlex RPR+. Negative BioPlex RPR results were observed in samples with reactive BD RPR: 10 (91%) of 11 for BD RPR 1:1, 13 (65%) of 20 for BD RPR 1:2, 6 (35%) of 17 for BD RPR 1:4, and 1 (7%) of 14 for BD RPR 1:8. The discordant samples were predominantly from patients with high-risk of syphilis reinfection and included 9 patients with an early reinfection.
Conclusions: Our results demonstrated that BioPlex Syphilis Total and Architect Syphilis TP performed similarly. The BioPlex RPR missed a small number of early syphilis reinfections, and its implementation should depend on the patient population that the laboratory serves.