Site qualification and clinical interpretation standards for 99mTc-SPECT perfusion imaging in a multi-center study of MITNEC (Medical Imaging Trials Network of Canada)

J Nucl Cardiol. 2021 Dec;28(6):2712-2725. doi: 10.1007/s12350-020-02100-9. Epub 2020 Mar 17.

Abstract

Background: Qualification and interpretation standards are essential for establishing 99mTc-SPECT MPI accuracy vs. alternative modalities.

Methods: Rest-stress 99mTc-SPECT phantom scans were acquired on 35 cameras. LV defects were quantified with summed stress (SSS) and difference scores (SDS) at 2 core labs. SDS ≥ 2 in the right coronary artery (RCA) was the qualifying standard. Twenty rest (R)-stress (S) patient images were acquired on qualified cameras and interpreted by core labs. Global scoring differences > 3 between labs or discordant clinical interpretations underwent review. Scoring, interpretation, image quality, and diagnostic parameter agreement were assessed.

Results: Phantom scans: visual scoring confirmed RCA-ischemia on all cameras. Regional SSS, SDS agreement was moderate to very good: ICC-r = 0.57, 0.84. Patient scans: 90% of global SSS, 85% of SDS differences were ≤ 3. Regional SSS, SDS agreement: ICC-r = 0.87, 0.86, and global abnormal (SSS ≥ 4) and ischemic (SDS ≥ 2) interpretation: ICC-r = 0.90 were excellent. Clinical interpretation agreement was 100% following review. Image quality agreement was 70%. Automated metrics also agreed: ischemic total perfusion deficit ICC-r = 0.75, reversible perfusion defect, transient ischemic dilation, and S-R LV ejection fraction ICC-r ≥ 0.90.

Conclusion: Quantitative scoring and interpretation of scans were highly repeatable with site qualification and clinical interpretation standardization, indicating that dual-core lab interpretation is appropriate to determine 99mTc-SPECT MPI accuracy.

Keywords: Ischemia quantification; Non-invasive myocardial perfusion imaging; Single-photon emission computed tomography; Standardization; Technetium-99m.

Publication types

  • Multicenter Study

MeSH terms

  • Aged
  • Aged, 80 and over
  • Canada
  • Female
  • Humans
  • Male
  • Middle Aged
  • Myocardial Perfusion Imaging / methods*
  • Phantoms, Imaging*
  • Technetium*
  • Tomography, Emission-Computed, Single-Photon / standards*

Substances

  • Technetium