Removal of complex aberrations at millisecond time scales over millimeters in distance in multiphoton laser scanning microscopy limits the total spatiotemporal imaging throughput for deep tissue imaging. Using a single low resolution deformable mirror and time multiplexing (TM) adaptive optics, we demonstrate video rate aberration correction (5 ms update rate for a single wavefront mask) for a complex heterogeneous distribution of refractive index differences through a depth of up to 1.1 mm and an extended imaging FOV of up to 0.8 mm, with up to 167% recovery of fluorescence intensity 335 µm from the center of the FOV. The proposed approach, termed raster adaptive optics (RAO), integrates image-based aberration retrieval and video rate removal of arbitrarily defined regions of dominant, spatially varied wavefronts. The extended FOV was achieved by demonstrating rapid recovery of up to 50 distinct wavefront masks at 500 ms update rates that increased imaging throughput by 2.3-fold. Because RAO only requires a single deformable mirror with image-based aberration retrieval, it can be directly implemented on a standard laser scanning multiphoton microscope.
© 2020 Optical Society of America under the terms of the OSA Open Access Publishing Agreement.