Background: Continuous kidney replacement therapy (CKRT) is a common modality for treatment of severe acute kidney injury (AKI) in children. Adult technologies routinely utilized to provide this therapy have a large extracorporeal volume. The Prismaflex™ HF20 filter set has a relatively low extracorporeal blood volume of 60 mL, which provides technological benefit for smaller children compared with current filter sets available in the USA.
Methods: We conducted a multicenter, open-label single group study to evaluate whether the Prismaflex™ HF20 filter set delivers efficacious and safe CKRT to support patients with AKI, fluid overload, or both in pediatric patients weighing ≥ 8 to 20 kg.
Results: Twenty-three patients were enrolled between April 24, 2016 and April 8, 2018. The mean reduction in blood urea nitrogen from baseline to 24 h was 58.12 ± 20.08% (95% CI, - 68.45 and - 47.79 (p = 0.0008)). Median cumulative normalized effluent rate at 24 h was 60.8 mL/kg/h (25.9, 83.7). None of the patients participating in the study suffered a serious adverse event; thus, no obvious safety concerns were noted.
Conclusions: We suggest that the Prismaflex HF20™ filter set used in conjunction with the Prismaflex™ System Software Version 7.10 or 7.20 is a suitable alternative to larger filter sets for use in pediatric patients weighing less than 20 kg. Graphical abstract.
Keywords: Acute kidney injury; CKRT; HF20; Kidney replacement therapy; Pediatric dialysis; Prismaflex.